Myocardium selective densitometric perfusion assessment after acute myocardial infarction

Tamás Ungi

a

, Imre Ungi

b

, Zsuzsanna Jónás

b

, Viktor Sasi

b

, András Lassó

c

, Zsolt Zimmermann

b

, Tamás Forster

b

, András Palkó

a

, Attila Nemes

b,

aDepartment of Radiology, Medical Faculty, Albert Szent-Györgyi Clinical Centre, University of Szeged, Szeged, Hungary

bSecond Department of Medicine and Cardiology Centre, Medical Faculty, Albert Szent-Györgyi Clinical Centre, University of Szeged, Szeged, Hungary

cGeneral Electric Healthcare, Budaörs, Hungary

Received 11 August 2008; received in revised form 14 October 2008; accepted 23 October 2008

Abstract Background:Myocardial perfusion is an important prognostic factor after recanalisation in acute myocardial infarction patients. We present a computerized, densitometric measurement method to assess myocardial perfusion on phase-matched digitally subtracted coronary angiograms.

Methods and materials: Quantitative myocardial perfusion was assessed by the Gmax/Tmax

parameter of the time–density curves (TDCs) in infarct-related myocardial regions on X-ray coronary angiograms. Arteries were masked out from regions of measurement. This novel method has been compared with enzymatic infarct size, ST-segment resolution, and ejection fraction after successful revascularization of 62 patients with acute myocardial infarction.

Results: Significant correlations were found between Gmax/Tmax and enzymatic infarct size (R=−0.445, Pb.001), ST-segment resolution (R=0.364, P=.004), and ejection fraction (R=0.278, P=.029). Bland and Altman plot ofGmax/Tmaxreveals good interobserver agreement.

Conclusions:Gmax/Tmax of the TDC measured in the infarct-related myocardial area is a reliable parameter to assess clinical indicators of myocardial reperfusion. Therefore, results suggest that it could be used to immediately assess the success of recanalisation at the tissue perfusion level during coronary intervention, and as an objective end point in clinical trials of new interventional devices and drugs.

© 2009 Elsevier Inc. All rights reserved.

Keywords: Acute myocardial infarction; Cineangiography; Myocardial viability

1. Introduction

The primary objective of treatment for acute myocardial infarction (AMI) is to restore normal blood flow in the epicardial infarct-related artery and to obtain reperfusion of the myocardium at risk. Assessment of myocardial perfusion on coronary angiograms is performed by visual estimation in the current clinical practice. Two visual grading scales proved to be informative in assessing the viability of the infarct-related myocardium: Myocardial Blush Grade[1]and TIMI Myocardial Perfusion Grade[2]. Both grades have four

This study was partially supported by the Regional Cooperative Research Center of Life and Material Sciences of the University of Szeged together with industrial partner General Electric Healthcare Hungary.

Dr. Tamás Ungi holds a Ph.D. scholarship from the University of Szeged.

Corresponding author. Medical Faculty, Second Department of Medicine and Cardiology Centre, University of Szeged, Korányi fasor 6, P.O. Box 427, H-6720 Szeged, Hungary. Tel.: +36 62 545220; fax: +36 62 544568.

E-mail address:nemes@in2nd.szote.u-szeged.hu(A. Nemes).

1553-8389/09/$see front matter © 2009 Elsevier Inc. All rights reserved.

doi:10.1016/j.carrev.2008.10.001

(2)

levels: 0 and 1 representing no or minimal contrast signal in the infarct related myocardium and 2 and 3 representing impaired and normal states. The interobserver and intraob- server variabilities associated with subjective angiographic assessments are limitations of these visual grades.

Methods for computerized videodensitometric analysis of digital subtraction coronary angiograms have been reported to be under study for a long time both in animal models and in human subjects. The primary goal of these studies was to develop an operator-independent and quantitative way of myocardial perfusion assessment based on X-ray coronary angiograms, which is the only imaging modality widely available during coronary interventions. Pijls et al. [3]

demonstrated in a canine model that mean transit time calculated by videodensitometry can be used to assess myocardial perfusion. Haude et al. [4] also proved an excellent correlation of a similar method with myocardial perfusion using colored microspheres. More recent reports proved that even volumetric blood flow measurements are feasible by densitometry[5,6].

Despite the promising results, none of these methods has yet been shifted to the clinical practice or became a standard end point of clinical trials in place of the four-graded visual assessment. Clinical environment, especially in case of AMI patients, does not allow the direct application of methods developed for experimental studies. Korosoglou et al. [7]

recently published the results of a study of AMI patients proving that the computerized method for myocardial perfusion assessment is clinically feasible. Their results suggest that this low-cost measurement provides incremen- tal prognostic value compared to the semi-quantitative visual assessment. However, the software used is not yet available and also has other limitations[8], e.g., coronary arteries should be excluded from the region of measurement, which severely limits the myocardial area where the method is applicable.

The objective of the present study was to evaluate regional myocardial perfusion assessed by a novel compu- terized videodensitometric method, and its relation to electrocardiographic, echocardiographic, and enzymatic indicators of myocardial damage in AMI patients after coronary intervention.

2. Materials and methods 2.1. Study populations

The study comprised 62 patients with AMI, who under- went primary percutaneous coronary intervention (PCI) at the Invasive Cardiology Division of the University of Szeged. Patients with the following inclusion criteria were enrolled into the present study: (1) acute ST-elevation on 12-lead ECG; (2) pain-to-balloon time b12 h; (3) total occlusion of the proximal segment in one of the three main coronary arteries; and (4) ability of the patient to cooperate.

Demographic and clinical data of the patients are shown in Table 1. Patients who were unconscious or showed signs of cardiogenic shock or had visible collateral circulation in the infarct-related myocardial region were excluded from the study. Informed consent was obtained from each patient and the study protocol conformed to the ethical guidelines of the 1975 Declaration of Helsinki, as reflected in a priori approval by the institution's human research committee.

2.2. Technical features of coronary angiography

Angiograms for densitometric analysis were recorded in a way that phase-matched digital subtraction angiography (DSA) can be performed on them. This required the following criteria: (1) motion of the patient or the table should be avoided, (2) patient should hold their breath during the period of the recording, (3) one contrast-free heart cycle should be recorded before injection of contrast material, (4) field of view is to be set to contain the whole supplied area of the vessel of interest. All coronary angiograms met these criteria.

Patients who were not able to hold their breath during the period of the recording were excluded from the study, which occurred in 11 (15%) of 73 cases. Projections were chosen to minimize the superpositioning of non–infarct-related myo- cardium and edge of the diaphragm, which usually gives motion artifacts on DSA images. Left anterior descending coronary artery and left circumflex artery were recorded in lateral (LAO 90°), while the right coronary artery was recorded in antero–posterior projection. The same nonionic contrast material (Visipaque, 320 mg/ml iodine) was used for all angiograms injected by a manual injector; contrast quantity was 6.84±0.97 ml (mean±S.D.); injection rate was 3.04±0.34 ml/s. Angiograms were recorded on an Innova 2000 system (GE Healthcare, Chalfont St. Giles, Buck- inghamshire, UK); images were stored in 512×512-size, 8-bit, grayscale, uncompressed format.

2.3. Echocardiographic, electrocardiographic, and enzymatic measurements

Echocardiographic measurements were performed 3 days after the primary PCI to assess left ventricular ejection fraction

Table 1

Clinical features of the patients Clinical parameters

Sample size (n) 62

Male gender (%) 43 (70)

Age (years±S.D.) 60±6

Diabetes mellitus (%) 6 (10)

Hypertension (%) 45 (73)

Smoking (%) 37 (60)

Infarct-related artery

Left anterior descending coronary artery (%) 31 (50)

Left circumflex coronary artery (%) 11 (17)

Right coronary artery (%) 20 (33)

(3)

(LV-EF). Twelve-lead electrocardiograms were recorded at the beginning of PCI and 90 min later. ST-resolution was defined as a decrease of ST-segment at 90 min compared to the first measurement in the lead with highest ST-segment elevation, expressed as percentage of initial ST elevation. Blood creatine kinase (CK) enzyme levels were measured 6, 12, 24, and 48 h after the PCI. These four measurements were summed up to assess total enzymatic infarct size.

2.4. Densitometric analysis

Phase-matched DSA angiograms were recorded with constant contrast, brightness, and stabilized acquisition para- meters of the X-ray imaging system. The computerized method for myocardial perfusion assessment was based on the analysis of time–density curves (TDCs) measured over the infarct- related myocardial region of interest (ROI). Polygonal-shaped ROIs were selected by an experienced interventional cardiol- ogist and covered the whole myocardial area at risk. Vesselness probability maps were calculated for each image frame according to Frangi et al. [9]. Typical ROI selections and

related TDCs are shown inFig. 1. Time–density curves were calculated as average pixel value in the ROI, excluding pixels having a vesselness probability valueN.08. Frequencies higher than 0.6 Hz have been removed from the TDC to eliminate artifacts from cyclic heart contractions, and noise in image acquisition and digital subtraction. Myocardial perfusion has a single-wave density signal around 0.1 Hz; therefore it is still present on the filtered curve. Frequency filtering was performed using the Matlab 7.0 mathematical software.

Maximal density of the TDC (Gmax) and time to reach maximal density (Tmax) were measured on the filtered curve. To get more reliable results and to assess interobserver variability of this new measurement method, two independent physicians were required to select the myocardial area at risk and draw a polygonal ROI. TwoGmax/Tmaxvalues calculated with the two different ROIs were averaged, and these results were used to analyze correlation with other clinical parameters. Cardiolo- gists were blinded to all other clinical data.

2.5. Statistical methods

All statistical tests were performed with MedCalc software (MedCalc Software, Mariakerke, Belgium). A

Fig. 1. Upper panel shows a frame of the DSA angiogram of the left coronary artery. White dashed line surrounds the region of densitometric measurement.

Lower panel shows the videodensitometric measurement result. Thinner curve represents the original average density; the thicker curve is computed by eliminating frequencies higher than 0.6 Hz from the original curve.

Fig. 2. Upper panel shows the BlandAltman plot of the difference between paired measurements by Observer 1 and Observer 2 for interobserver comparison ofGmax/Tmax. Scatter plot of the same paired measurement results is shown in the lower panel.

(4)

value ofPb.05 was considered to be statistically significant.

Correlation of Gmax/Tmax with clinical parameters was assessed by Pearson correlation coefficient. Interobserver variability was analyzed by the Bland–Altman method.

3. Results

All patients underwent a successful recanalisation of the occluded vessel and achievedb50% residual stenosis within 12 h from the onset of symptoms. Plotting the Gmax/Tmax

values on Bland–Altman plot and scatter diagram obtained by two independent observers reveals a reliable interobserver agreement (Fig. 2) of our method.

Enzymatic infarct size as expressed by sum of CK release had a significant negative correlation (R=−0.445, Pb.001) withGmax/Tmax. Scatter plot and regression line along with 95% predictive boundary are shown inFig. 3. Additionally, a positive significant (R=0.364, P=.004) correlation was found between Gmax/Tmax and ST-segment resolution as % decrease of initial ST elevation (Fig. 4). A milder, but still significant relationship was found between Gmax/Tmax and echocardiographic LV-EF measured 3 days after PCI (R=0.278,P=.029) (Fig. 5).

4. Discussion

Results of the present study show that the demonstrated computerized videodensitometric method for regional myo- cardial perfusion assessment can be used in clinical circumstances to get immediate information on the viability and functionality of the revascularised myocardium. More- over, measurements correlate with clinical indicators of myocardial damage suggesting it as an end point for studies on the success of recanalisation in AMI.

There is an abundance of evidence from several techniques, such as intracoronary contrast echo [10,11], magnetic resonance imaging (MRI) [12], and radionuclide studies [13], showing that many patients have inadequate flow at myocardial tissue level despite a reopened epicardial coronary artery. Low microcirculatory perfusion after angioplasty results in larger enzymatic infarct size, lower residual ejection fraction, higher mortality, and higher incidence of major adverse cardiac events in a long-term follow-up[14]. In daily practice, coronary angiography is the most used technique during recanalisation procedure to study the effectiveness of reperfusion therapy. Assessment of microcirculatory perfusion in the myocardium is therefore necessary even if the recanalisation of occluded epicardial artery is successful (TIMI 3 flow grade). The demonstrated

Fig. 4. Correlation betweenGmax/Tmax and ST-resolution measured as a relative decrease in ST elevation between PCI and 90 min after coronary intervention. There is a significant positive correlation between these two parameters. Thick line indicates regression line; thin lines indicate 95% prediction.

Fig. 5. Correlation between Gmax/Tmax and echocardiographic left ventricular ejection fraction 3 days after coronary intervention. Thick line indicates regression line; thin lines indicate 95% prediction.

Fig. 3. Correlation betweenGmax/Tmaxand enzymatic infarct size as sum of four creatine kinase blood levels (measured 6, 12, 24, and 48 h after coronary intervention). Thick line indicates regression line; thin lines indicate 95% prediction.

(5)

method provides direct information on the flow of contrast agent through myocardial areas, where vessel patency cannot be visually assessed due to small vessel diameters and limited imaging resolution.

In most previous studies [3,7], myocardial ROI was smaller than in our method. We have chosen to include the whole distribution bed of the infarct-related vessel in the ROI because the distribution of contrast material in the myocar- dium is generally not homogeneous. During our previous investigations, smaller ROIs resulted in greater interobserver variability, because measurement values varied much if the ROI was placed at different locations inside the infarct- related area by different observers. To our knowledge, the present study is the first to combine vessel masking with perfusion assessment on X-ray coronary angiograms. This allowed us to define ROIs including the whole myocardial area at risk, while arterial contrast density still did not interfere with myocardial density signal analysis. Over- coming the problem of positioning small ROIs in the myocardial area, we have reached an almost observer- independent densitometric method, as reflected by our inter- observer agreement (R2=0.97).

In most animal studies, the method of TDC analysis involves fitting of an ideal curve to measurement data. Pijls et al. [3] used least squares minimization with gamma function. A two-compartment mathematical model was also proven to be useful for regional perfusion assessment in a canine model[15]and later in humans[16]. Haude et al.[4]

used a lognormal curve fitting method to compute densito- metric parameters of contrast perfusion. We have chosen a more simple curve smoothing method because multipara- meter curve fitting can be a source of numerical errors and algorithmical variability. Our previous investigations show that the parameters of the fitted curve depend strongly on the size of the window that we use for fitting on the original curve. In cases of AMI, the wash-out phase is often missing from the acquisition, and motion artifacts cause temporary noise on the measurement curves. Gmax/Tmax calculation needs only two points of the filtered curve, and frequency filtering is enough to accurately locate these points; therefore this method is simple, robust, and can be applied in most of the cases.

4.1. Study limitations

As already mentioned, distribution of contrast material in infarct-related myocardium is not homogeneous. Division of this area and analysis of smaller regions have not been performed in the framework of this study. The problem with small ROI areas is that a static ROI was used during the whole image sequence, and cyclic motion of myocardium due to heart beating increases relatively to the size of ROI. A possible improvement of our method could be the motion tracking of myocardial regions. This could enable us to perform measurements in smaller regions and could give a more precise localisation of the myocardial injury.

Another limitation of our method is that it requires a meaningful cooperation of the patient while recording images for phase-matched DSA conversion. Patients with AMI are often unable to hold their breath during the period of image acquisition. During the enrollment period of this study, 11 (15%) of 73 patients could not be enrolled because of lack of cooperation. Since the breathing of the patient makes densitometric measurements extremely difficult, in such cases other myocardial perfusion assessment methods must be used; myocardial contrast echocardiography may be an alternative.

Contrast material quantity somewhat differed between image acquisitions (6.84±0.97 ml), and injection rate also had 11% standard deviation (compared to mean value) (3.04±0.34 ml/s). It was hypothesized that variability in total contrast quantity and injection rate did not affect the rise in the slope of the TDC, because contrast material completely filled the coronary arteries at the point of injection, in each image acquisition.

Measurement of CK release, LV-EF, and ST resolution are only indirect indicators of reperfusion. Validation of this novel videodensitometric method could be more valuable if it is compared to other imaging modalities that directly assess regional myocardial perfusion. A comparison with myocardial contrast echocardiography, MRI, or positron emission tomography measurements could provide stronger evidence or valuable information on this method. However, videodensitometry for myocardial perfusion assessment is strongly supported by previous animal experiments and human studies as well. The weak correlation compared to echocardiographic LV-EF may be explained by myocardial stunning that can still be present 3 days after the infarction.

Follow-up studies may reveal stronger correlation between Gmax/Tmaxand left ventricular function.

4.2. Clinical implications

A good quality DSA coronary angiography can be acquired in the majority of AMI patients. Based on the results of the present study, it can be stated that videodensi- tometric measurements on coronary angiograms can be developed and used in cardiac catheterization laboratories.

Videodensitometry proves to be a low-cost alternative to other imaging modalities like myocardial contrast echocar- diography or MRI to immediately assess success of PCI, but further validation studies are warranted.

References

[1] van't Hof AW, Liem A, Suryapranata H, Hoorntje JC, de Boer MJ, Zijlstra F. Angiographic assessment of myocardial reperfusion in patients treated with primary angioplasty for acute myocardial infarction: myocardial blush grade. Zwolle Myocardial Infarction Study Group. Circulation 1998;97:23026.

[2] Gibson CM, Cannon CP, Murphy SA, Ryan KA, Mesley R, Marble SJ, McCabe CH, Van De Werf F, Braunwald E. Relationship of TIMI

(6)

myocardial perfusion grade to mortality after administration of thrombolytic drugs. Circulation 2000;101:12530.

[3] Pijls NH, Uijen GJ, Hoevelaken A, Arts T, Aengevaeren WR, Bos HS, Fast JH, van Leeuwen KL, van der Werf T. Mean transit time for the assessment of myocardial perfusion by videodensitometry. Circulation 1990;81:133140.

[4] Haude M, Caspari G, Baumgart D, Ehring T, Schulz R, Roth T, Koch L, Erbel R, Spiller P, Heusch G. X-ray densitometry for the measurement of regional myocardial perfusion. Basic Res Cardiol 2000;95:26170.

[5] Molloi S, Zhou Y, Kassab GS. Regional volumetric coronary blood flow measurement by digital angiography: in vivo validation. Acad Radiol 2004;11:75766.

[6] Wong JT, Ducote JL, Xu T, Hassanein MT, Molloi S. Automated technique for angiographic determination of coronary blood flow and lumen volume. Acad Radiol 2006;13:18694.

[7] Korosoglou G, Haars A, Michael G, Erbacher M, Hardt S, Giannitsis E, Kurz K, Franz-Josef N, Dickhaus H, Katus HA, Kuecherer H.

Quantitative evaluation of myocardial blush to assess tissue level reperfusion in patients with acute ST-elevation myocardial infarction:

incremental prognostic value compared with visual assessment. Am Heart J 2007;153:61220.

[8] Malsch U, Dickhaus H, Kuecherer H. Quantitative Analyse von Koronarangiographischen Bildfolgen zur Bestimmung der Myokard- perfusion. Bildverarbeitung fur die Medizin 2003:815.

[9] Frangi AF, Niessen WJ, Vincken KL, Viergever MA. Multiscale vessel enhancement filtering. Lect Notes Comput Sci 1998;1496:1307.

[10] Ito H, Tomooka T, Sakai N, Yu H, Higashino Y, Fujii K, Masuyama T, Kitabatake A, Minamino T. Lack of myocardial perfusion immediately

after successful thrombolysis. A predictor of poor recovery of left ventricular function in anterior myocardial infarction. Circulation 1992;85:1699705.

[11] Ito H, Maruyama A, Iwakura K, Takiuchi S, Masuyama T, Hori M, Higashino Y, Fujii K, Minamino T. Clinical implications of theno reflowphenomenon. A predictor of complications and left ventricular remodeling in reperfused anterior wall myocardial infarction. Circula- tion 1996;93:2238.

[12] Wu KC, Zerhouni EA, Judd RM, Lugo-Olivieri CH, Barouch LA, Schulman SP, Blumenthal RS, Lima JA. Prognostic significance of microvascular obstruction by magnetic resonance imaging in patients with acute myocardial infarction. Circulation 1998;97:76572.

[13] Kondo M, Nakano A, Saito D, Shimono Y. Assessment of

microvascular no-reflow phenomenon using technetium-99m macroaggregated albumin scintigraphy in patients with acute myo- cardial infarction. J Am Coll Cardiol 1998;32:898903.

[14] Henriques JP, Zijlstra F, van't Hof AW, de Boer MJ, Dambrink JH, Gosselink M, Hoorntje JC, Suryapranata H. Angiographic assessment of reperfusion in acute myocardial infarction by myocardial blush grade. Circulation 2003;107:2115–9.

[15] Eigler NL, Schühlen H, Whiting JS, Pfaff JM, Zeiher A, Gu S. Digital angiographic impulse response analysis of regional myocardial perfusion. Estimation of coronary flow, flow reserve, and distribution volume by compartmental transit time measurement in a canine model.

Circ Res 1991;68:87080.

[16] Schühlen H, Eigler NL, Zeiher AM, Rombach MM, Whiting JS.

Digital angiographic assessment of the physiological changes to the regional microcirculation induced by successful coronary angioplasty.

Circulation 1994;90:16371.

Hivatkozások

KAPCSOLÓDÓ DOKUMENTUMOK

In the present study, our aim was to assess the endoscopic activity and the rate of mucosal healing after a one-year period of biological therapy and to evaluate how the

Detection of early therapeutic changes of the antiangiogenic drug, bevacizumab, using a dual agent MR imaging method (ferumoxytol for perfusion assessment and

Although the induction of acute myocardial infarction in nondiabetic rats significantly decreased systolic performance (ejection

In a study intended to test the potentiating effect of 2 different cardioprotective agents, the NACIAM (N-acetylcysteine in Acute Myocardial Infarction) trial (56) examined the

As such, in order to prevent heart failure and improve clinical outcomes in patients presenting with an acute ST-segment elevation myocardial infarction and patients undergoing

Isoproterenol (85 mg/kg) was injected subcutaneously 2 days after the first treatment at an interval of 24 hours for 2 days to produce myocardial infarction. After 17

Regional myocardial ischaemia was induced in the rodent model by ligation of the left anterior descending (LAD) coronary artery. Global myocardial ischaemia was

Figure 5. Experimental protocol to test remote ischemic conditioning in an in vivo porcine model of acute myocardial infarction. Isch – ischemia only, IPreC – ischemic