• Nem Talált Eredményt

Trends in Cardiovascular Medicine ARTICLE IN PRESS

N/A
N/A
Protected

Academic year: 2022

Ossza meg "Trends in Cardiovascular Medicine ARTICLE IN PRESS"

Copied!
6
0
0

Teljes szövegt

(1)

ContentslistsavailableatScienceDirect

Trends in Cardiovascular Medicine

journalhomepage:www.elsevier.com/locate/tcm

Rationale and feasibility of the atrioventricular single-lead ICD systems with a floating atrial dipole (DX) in clinical practice

Mate Vamos

a,

, Marianna Nemeth

b

, Tibor Balazs

c

, Endre Zima

d,1

, Gabor Zoltan Duray

e,1

aDepartment of Internal Medicine, Cardiac Electrophysiology Division, University of Szeged, Semmelweis u. 8., 6725 Szeged, Szeged, Hungary

bHeart Institute, University of Pécs, Pécs, Hungary

cBiotronik Hungary Ltd., Budapest, Hungary

dHeart and Vascular Center, Semmelweis University, Budapest, Hungary

eDepartment of Cardiology, Medical Centre, Hungarian Defense Forces, Budapest, Hungary

a rt i c l e i nf o

Keywords:

ICD

Implantable cardioverter defibrillator Single-lead ICD

VDD DX

Two-lead CRT-D CRT-DX

a b s t r a c t

Cardiac implantableelectronic devices establish proper therapy for the prevention of sudden cardiac death,significantlyreducingthe morbidity and mortality ofpatientswith arrhythmiasand heartfail- ure.Itiswell-knownthatthenumberofelectrodesincreasestheriskofcomplications.Topreservethe benefitofatrialsensingwithouttheneedtoimplantanadditionallead,asingle-leadICDsystemwitha floatingatrialdipole(DXICDlead)hasbeendeveloped.Besidesallofthepotentialbenefits,thenecessity ofareliableandstableatrialsensingviathefloatingdipolecouldbethemainconcernagainsttheuseof thisleadtype.InthecurrentgenerationofDXdevices,thespeciallyfilteredatrialsignalseemstobehigh enoughandstableovertime,whichiscrucialintheearlydetectionofatrialarrhythmias,discrimination betweendifferentformsoftachycardiasinordertopreventinappropriateICDtherapy,andachievingan optimalatrioventricularandinterventricularsynchronyinpatientswithatwo-leadCRT-DXsystem.

Thepresentreviewsummarizesthebenefitsandpotential drawbacksofthe DXICDsystemsbasedon theavailableliterature,furthermore,proposesanevidence-basedalgorithmofICDtypeselection.

© 2021TheAuthor(s).PublishedbyElsevierInc.

ThisisanopenaccessarticleundertheCCBY-NC-NDlicense (http://creativecommons.org/licenses/by-nc-nd/4.0/)

Introduction

Cardiacimplantableelectronicdevices(CIEDs)significantly im- prove the morbidity andmortality of patientssuffering from ar- rhythmiasandheartfailure.Althoughthesesystemshavebecome more andmore safeandeffectivedueto significanttechnicalad- vancesofthelastdecades,asubstantialriskforserious complica- tions relates to thetransvenously implantedleads. It isalso well known that the number of leads increases not only the time of surgeryandfluoroscopybutalsotheriskofcomplications,particu- larlyofthrombosis,infection,ortheneedforrepeatedsurgery[1– 5].Differenttechnologies,utilizingextravascularleads,suchasthe subcutaneousICD, havebeendevelopedtoprovidethepossibility to convertlife-threatening ventriculararrhythmias whileavoiding complications relatedto theintravascular lead[6].Thesesystems are, however, currently limited to patients without the need for

Corresponding author.

E-mail address: vamos.mate@gmail.com (M. Vamos).

1 These authors contributed equally as last authors.

pacingtherapy forbradycardiasupport,cardiacresynchronization, andantitachycardiapacing[7].Itseemsreasonabletoimplantonly thoseleadsthatareultimatelyrequired.

Asingle-leadICDsystemwithafloatingatrialdipole(DXtech- nology)wasdevelopedtopreservethebenefitofsensingatrialac- tivitywithouttheneedtoimplant anadditionalatriallead[8].In thepresentpaper,wesummarizethepotentialbenefitsanddraw- backsofDXtechnologybasedontheavailableevidenceinthelit- erature.

DescriptionoftheDXtechnologyinimplantabledefibrillators

DX technology and the term “DX” specify the Diagnostic eXtension capability of the DX ICD lead (DF-1 and DF-4 vari- ants),whichisaventricularICDleadwithanaddedfloatingatrial dipolein order to provide atrial sensing insingle-chamber ICDs.

Thetechnology wasimplemented inBiotronikICDs(Biotronik, SE

& Co., Berlin, Germany) in the early 2000s under the name of A+system, anditwasrelaunched asDX technologyinEuropein 2011andin theUS in2013[9].The historyofit witha detailed

https://doi.org/10.1016/j.tcm.2021.01.003

1050-1738/© 2021 The Author(s). Published by Elsevier Inc. This is an open access article under the CC BY-NC-ND license ( http://creativecommons.org/licenses/by-nc-nd/4.0/ )

Pleasecitethisarticleas:M.Vamos,M.Nemeth,T.Balazsetal.,Rationaleandfeasibilityoftheatrioventricularsingle-leadICDsystems

(2)

Fig. 1. (Graphical abstract) Overview of the different DX ICD systems and their potential benefits in different clinical situations.

technicaldescriptionhasbeenpublishedbefore[8],butsome im- provementshavebeenmadesincethattimeanda2-leadCRTICD systemhasalsobeenintroduced.Herein,wepredominantlyfocus on thelast generation ofthesesystems and, thereforethetermi- nology“DX” willbeusedingeneral.TheDXICDsystemreferstoa

“DX” single-leadICDwithatrialdiagnosticsbutwithoutatrialpac- ing capability,whichconsistsofaDXICD generatorandaDXICD lead (Fig.1).TheCRT-DXsystemisa2-leadCRT-Dsystem, which maintains the AV-synchronous resynchronizationtherapy via two leads. ItconsistsofaCRT-DXdevice,aDXICD lead,andcoronary sinuslead(Figs.1and2).

ThecurrentlyavailableDXICDandCRT-DXsystemshaveasta- ble and reliable atrial sensing, which is based on the optimized atrial dipole spacing, a specifically designed input stage for the atrial sensing, a pre-amplifier, which progressively increases the atrialgainup tofourfold,andawiderbandpassforthefrequency range oftheatrial channel,respectively. Improvements oftheDX technology provide flexibility of lead positioning in the atrium witha reducedrisk offar-fieldoversensingfromtheventricle.In addition, the loss of atrial signal results in automatic switching fromSMARTdetectiontosingle-chamberdiscrimination.

PotentialbenefitsandrisksoftheDXICDlead Clinicalbenefitsrelatedtobradycardiapacing

In dailyclinical practice,only 40% ofdual-chamber ICD recip- ients fulfill an indication for dual-chamber pacing [9], however more complications are observed in dual-chamber ICD recipients [1,4,10,11]. Notably, dislodgement of the atrial lead occurs more often compared to the rightventricular leads with an incidence

Fig. 2. Chest X-ray of a patient suffering from dilatative cardiomyopathy after an upgrade from DX ICD to CRT-DX with a quadripolar coronary sinus lead.

ofup to 1.9% [12].A needfor atrialpacing in thecase ofsymp- tomaticsicksinussyndromeseemstobeaclearindicationforim- plantinganatrialleadand, therefore,inthesecases,DXICDleads arenot an option.Inselected patientsaftermyocardialinfarction and/orwhoare sufferingfromheartfailureandclinicallyrelevant sinusbradycardialimitingbeta-blockertherapyuptitration,theim- plantationofanatrialleadmaybeconsidered.Unnecessaryatrial

(3)

stimulationis,however,potentiallydangerousinthatatrialfibrilla- tionispromoted,asobservedinseveralstudies[13,14].Therefore, an additionalatrialleadshouldbe avoidedinsinglechamber ICD recipients withoutsinusnodedisease.Clinicians tendtochoosea dual-chamber system over VVIICD for patientswho still do not fulfill an absolute indication for bradypacing but already have a mild AV orventricularconductionproblem(i.e. first-degreeheart block, rightorleftbundle branchblock,widerQRSduration),and maydevelophigherdegreeconductionabnormalitieslater[10].In thecaseofconcomitantICDindicationandmildAVorventricular conductionproblems,butwithoutan absoluteindicationforatrial pacing, a DX ICD lead maybe an optimal choice. In the case of a left bundle branch block or an expectedhigh rateof ventricu- larpacinginthefuture,primarycardiacresynchronizationtherapy should be strongly considered toprevent a CRT upgrade lateron [15].

Cardiacresynchronizationvia2leads(CRT-DXsystems)

A special clinical question is whether CRT can be delivered safely with 2 leads (CRT-DX system), since the reduction in the numberofthe leadsismorerelevantinthesecomplexmultilead systems.Stable atrialsensingiscrucialinCRTinordertoachieve optimalatrioventricular(AV)andinterventricular(VV)timing.

The role of atrial stimulation in CRT recipients without sinus nodediseaseorsymptomaticchronotropicincompetenceiscontro- versial.InapropensitymatchedstudyofCRT-Drecipientsfollowed by theLATITUDE remotemonitoringsystem,rate-responsivedual- chamber(DDDR)programmingwasassociatedwithsignificantim- provementinsurvivalbutonlyforaselectedhigh-riskpopulation identified by heart ratescore, a novel parameter of chronotropic incompetence[16].Onthecontrary,thecompositeendpointofall- cause mortality, heart failure events, NYHA functional class, and patientglobalself-assessmentinamulticenter,3-arm,randomized studyofCRTrecipientswerenotimprovedbyatrialsupportpacing (DDD-70 or DDDR-40) compared to atrial tracking (DDD-40)[17].

Notably, the 2012 EHRA/HRSexpert consensus statementon car- diacresynchronizationtherapyrecommendsusingVDD/DDDmode withabaserateof35–40bpmtoensurepermanentornearlyper- manentatrial sensingandto avoidtheconfounding influences of atrialsupport[18].

Regarding clinical experience with CRT-DX systems, there is still limited evidence. Biffi and colleagues compared the clinical and technical outcome of 25 CRT-D recipients with conventional 3-lead systems and 12 patients implanted with a CRT-DX [19]. Absolute indication for atrial stimulation was observed in 1.2%

of CRT recipients during the 2-year screening period and these patients were accordingly excluded from the study. Besides the classical evaluation of CRT response, chronotropic incompetence was also carefully investigated at baseline and during follow-up.

No difference between the 2 groups was observed in terms of NYHA classimprovement,LV reverseremodeling, peakcardiopul- monary performance, and the presence of chronotropic incom- petence at 12 months. Moreover, no patients developed a need for atrialstimulation at 3-year follow-upandatrial undersensing did not occur. In another publication, 120 subjects with CRT-DX fromtheSentusQP-ExtendedCRTEvaluationwithQuadripolarLeft Ventricular Leadspostapprovalstudy(Clinicaltrials.govIdentifier:

NCT02290028)wereselectedandmatchedwith120patientswith conventionalthree-leadCRT-Dsystems.Thisretrospectivesubanal- ysisdemonstratedthattheCRT-DXsystemcanprovidefewercom- plicationsdrivenprimarilybylackofRAleaddislodgements,while similarCRTresponseswereachieved[20].

To provide optimalCRT, the reliable detection of atrial fibril- lation (AF) andan adequate mode switch is crucial,independent from thetype ofatrial sensingelectrodes. InCRT recipients with

permanent AF, atrial lead is usually not implanted, however, a spontaneousreversion intosinusrhythmmayoccur inup to10%

[21].Thesepatientsmayhavea benefitfromaCRT-DXsystemas they can have an AV synchronous pacing without riskingan ad- ditional atrial lead implantation. CRT upgrading is burdened by complications and infection primarily owing to the lead number [15]and, therefore,upgrading aDX ICD device -withan already provenstableatrialsensing-seemstocarrylowerrisk(Fig.2).In theTHINGSregistry,oneoutofthe140ICDDXpatientsdeveloped anindicationforupgradingthedevicetoacardiacresynchroniza- tiontherapy defibrillator(CRT-D):the upgradingwassuccessfully completed usingthesame DXtechnology withatwo-lead CRT-D system[22].

Despitetheseencouragingobservationaldatarelatedtotheuti- lizationoftheCRT-DXsystem,randomizedcontrolledtrialsinclud- ing upgradefromDX ICDto CRT-DXare highlywarranted inthis field.

TachycardiadiscriminationtopreventinappropriateICDtherapy

Differentiationof supraventricularandventriculartachycardias are one of the most exciting but sometimes also the mostchal- lengingareasofcardiacelectrophysiology.Theavailabilityofatrial electrograms in addition to ventricular IEGM and far-field elec- trogramsimprovesthecorrectinterpretationofstoredarrhythmia events[9].Dual-chamberICDscouldprovidebetterautomaticdis- criminationbetweenSVTandVTbasedonatrioventriculardissoci- ation,althoughclinicalstudiesinvestigatingthisissuehavecontro- versialresults,especiallytheonesusedmorphologydiscriminators insingle-chamberdevices[23–25].

Nonetheless, according to the HRS/EHRA/APHRS/SOLAECE ex- pert consensus, improved SVT-VT discrimination should not be consideredasanindicationforadual-vs.single-chamberICD[26]. Moreover,theprogrammingofmorphologydiscriminators(i.e.Far- FieldMorphology,MorphMatch,RhythmID, andWavelet)are rec- ommendedinsingle-chamberICDsandtheprogrammingofdual- chamberalgorithms(plusmorphology,ifavailable)indeviceswith anatriallead.

The DX ICD system with atrial-sensing electrodes offers the same discrimination capability asdual-chamber ICDs (i.e.SMART algorithm) besides the single-chamber discrimination algorithms.

Insome early studies ofthe DXICD system, inappropriate thera- pies were observedin up to7.8–8.6% ofthe cases,dependingon the follow-up duration [27,28]. In a novel publication, however, thiswasonly0.83% duringthemean follow-upof1.3years [20], and in another cohort of 150 DX ICD recipients with a median follow-up of 12 months, there were no inappropriate ICD thera- piesobservedatall[29].Theobservedimprovementintherateof inappropriateICDtherapiesismostlikelyduetothemoreconser- vative programmingof major arrhythmia discriminators, highVT detection cut-off rate, and prolonged detection time. In a com- parative, prospective, single-center study of 212 consecutive pa- tients who underwent conventional (VVI) or DX single-lead ICD implantation, the incidenceof inappropriate ICD therapies in the DX-groupwassignificantlylowercomparedtotheVVI-group(1/77 [1%]vs.12/135[9%],p=0.028)[30].

Although dual chamber algorithms would be logically prefer- able,itisnotclearwhichdiscriminatorshouldbeempiricallypro- grammed as the first choice in thesedevices, since no head-to- headcomparisonbetweenmorphologyanddual-chamberdiscrim- inationmodesofBiotronikdevices(i.e.MorphMatchvs.SMART)is available.Fornote,dual-chamberdiscriminationmaybepreferred foratrial arrhythmiaswithslow andregular ventricular rhythms, as suggested in the THINGS registry [22]. Moreover, in the case ofinappropriate discrimination, changing the discrimination type

(4)

betweenmorphologyanddual-chamberalgorithmsshouldbecon- sidered.

Earlydetectionofatrialarrhythmias

Atrial fibrillation (AF) is associated with substantial mortality andmorbidity,includingstroke,heartfailure,cognitivedecline,de- pression, impaired quality oflife, andhospitalizations [31]. Atrial fibrillation increasesthe risk of stroke fivefold, buteven a short, subclinicaldevicedetectedatrialhighrateepisode(AHRE)isasso- ciated withanincreasedstrokerisk of0.8–1%peryearcompared with CIED patients without AHRE [32]. Early detection of silent atrialfibrillationmayhelpreducingstrokerisk,preventionofheart failureexacerbation,andavoidingatrialremodeling.

In theprospective, multicenterSENSEtrial,150 patientswith- outapriorhistoryofAFwereimplantedwithaDXICDsystemand the incidenceofAHREdetectionat12 monthswere comparedto age,sex,andleft ventricularejectionfractionmatchedsingle-and dual-chamberICDcohorts.TherateofAHREdetectionwassignifi- cantlyhigherintheDXcohortcomparedtothesingle-chamberco- hort (13%vs. 5.3%,p=0.026),butnotsignificantlydifferentcom- paredtothedual-chamber cohort(13%vs.13%, p= 1.00)[29].In themostrecentlypublishedprospective,observational,multicenter THINGS registry clinical outcomesofpatients implantedwithDX ICD withatrial sensingcapability (ICD DX group, n = 140) were compared to those implanted with conventional single-chamber ICD (ICD VR group, n = 236) [22]. The likelihood of atrial tach- yarrhythmia detectionwas almost fourfoldgreater in theDX ICD group(11.4%vs.3.6%within2years,adjustedHR3.85,95%CI1.58–

9.41,p =0.003). Arrhythmiasdiagnosed in thatwayoften ledto clinical interventions,mainlyby theinitiationoforal anticoagula- tiontherapy.

These data strongly support the utilization of a DX ICD over conventional single-chamber ICDs, especially for patients with a higherrisk forAF. InaproposalforICD selectionpublishedsome yearsago[33],adevicewithanalgorithmforAT/AFdetectionwas recommended for single chamber ICD candidates with a CHADS score≥2,representingpatientswithanelevatedrisk ofatrialar- rhythmias.Clinicalconditionsassociatedwithelevated strokerisk beyondtheonesincorporatedintotheCHADSscorearedescribed in detailby themostrecentESC guidelinesforthediagnosisand managementofAF[31].Accordingly,themajorityoftypicalICDre- cipientswouldbenefitfromtheAFmonitoringcapabilityoftheDX technology;howeverthereisnoscientificevidenceonhardclinical outcomesavailable.Combinationofremotemonitoringandthede- tectionofatrialarrhythmiashelpstostartanearly,appropriatean- ticoagulation therapyforpatientssufferingfromasymptomaticAF events.Theongoing randomizedDX-AFstudyisaimed todemon- strate that theDX ICD systemcombinedwith remotemonitoring willfacilitatetheadequaterecognitionofsub-clinicalAFandulti- mately stroke prevention withsimilar safety profilecompared to VVI-ICD(NCT03110627).

KnownandpotentialrisksoftheDXsystem

The previously discussed potential benefits rely on a stable atrialsignalviathefloatingdipole.Arandomizedstudyin2011re- portedthatarrhythmiadiscriminationwassimilarin249patients withan earlygenerationoftheDXICDsystem(A+ system)com- pared to dual-chamber ICD, however the percentage of patients with atrial over- or undersensing wassignificantly higherin the A+arm[34].Inanotherearlystudyof43patientswiththeLexos A+model,atrialundersensingwasneverdocumented[35].

The system has undergone a series of modifications to opti- mize atrial sensing. In the currentgeneration of the devices,av- erageatrial signalamplitudes rangefrom1.8to 8.7mVandhave

proventobestablethroughlong-termfollow-upsindifferentbody postures orlead positions,withvery few reportedcases ofatrial undersensing [27,28,36,37].In a recentstudyof 93patients with LinoxS DX ICD leadsfollowedby remote-monitoring system, the investigatorsconfirmedthattheleaddemonstratedexcellentfunc- tioningregardingthechronicstabilityofthecardiacsignals[36].In spiteofsignal stability,ahigh-rateofmisclassificationofarrhyth- miawereobservedthatwasprimarilycausedbyfar-fieldoversens- ingoftheventricularsignalintheatrialchannel.Apartfrom7.5%

inappropriateICDtherapy,inthreepatients,VTepisodeswereclas- sifiedassupraventricularandaccordinglyremaineduntreated.The reasonformisclassificationswasasinglebeat atrialundersensing duetoventricularblankingandthedeviceswitchedfromthedual- chamber tothesingle-chamberalgorithm (i.e.onset/stability).Al- thoughinthiswaytheriskofinappropriatetherapyistheoretically reduced,whenevertheswitchinthediscriminationalgorithmoc- curs duringan episodeof ventricular arrhythmia,the arrhythmia onsetcriterion is lost andthemalignant arrhythmiaremains un- diagnosed, as pointed out by Safak et al. [36]. A similar caseof atrialundersensingbytheDX ICDleadresultingindetectionfail- ure ofa slow ventricular tachycardiawas reportedwitha newer generationofDXICD(BiotronikItrevia7VR-TDX)[38].Intheset- tingoflossofatrialsensing, morphologydiscriminationmaybe a moreeffectiveSVT/VTdiscriminationstrategy,butthecombination ofmorphology discriminationwiththe SMARTalgorithm isnot a programmableoptionatthepresenttime.

ThemostrecentdatafromtheTHINGSregistryshow98.6%sen- sitivity of appropriate atrial signal detection (median P wave at baseline5.5mV,atthe2-yearfollow-up5.5mV).Intermittentin- adequate atrialsensingwas observedin two patients(1.4%)with a mean atrial signal amplitude of 0.4 mV [22]. Similarly, in the SENSEtrial,themeansensedatrialamplitudewiththeDXICDlead was8.0±5.0mV atimplant and7.3± 4.8mV atthe12-month follow-up[29].Moreover,theaccuracyofAHREdetectionintheDX cohortwascomparabletothatofthedual-chambercohort(13%of AHREdetectionsintheDXcohortand9%inthedual-chamberco- hort were false positives).Individual casesof electromagneticin- terferencecausedby signal amplificationhavealsobeenreported [39,40].

Regarding operativeparameters, neitherprocedure/fluoroscopy durationsnorperioperative complication ratesseem tobe higher thantheonesobservedduringconventionalVVIICDimplantations [29,30].DefibrillatorswithDXtechnologyalreadyhavetheoption to be remotely monitored, which provide the possibility of early detectionandanalysisofthepotentialrisksdetailedabove.

There are also some open questions aspotential cons for the utilizationoftheDXsystem,suchaslong-termperformanceofthe leadorlead extraction.There isactuallyno researchavailable re- latedtocost-effectivenessand,therefore,anystatementonthisis- sueisalsonotpossible.

ProposalforICDselection

BasedonscientificevidenceoftheDXICDandCRT-DXsystems, assummarizedinthisreview,thecurrentEuropeanandAmerican guidelinesregardingcardiacpacing,resynchronizationtherapy,and therecommendedscreeningofatrialfibrillation [26,31,41–43],we suggestconsidering thefollowingclinicalquestionsinthecaseof aplannedICDimplantation.

§ Isthereaneedfortheimplantationofaleftventricularlead?

• Symptomaticheartfailure,LVEF≤35%andLBBBQRS>130ms

• Symptomatic heart failure, LVEF ≤35% and non-LBBB QRS >

150ms

• NarrowQRSbutLVEF36–50% andanticipatedrequirementfor significantventricularpacing[44]

(5)

ICD indication (primary or secondary prophylactic)

Indication for CRT 1,2

Indication for atrial stimulation 3,4

CRT-DX 11 Yes

Clinical need for AV-sequential pacing 5

OR Atrial monitoring 6

DX ICD 10 DDD ICD CRT-D

No No

No

Yes

Yes

Indication for CRT

• Symptomatic HF AND LVEF ≤ 35% WITH LBBB QRS ≥ 130ms OR non-LBBB QRS ≥ 150ms 1

• Narrow QRS BUT LVEF 36-50% AND anticipated requirement for significant ventricular pacing 2 VVI ICD

No

Yes

Indication for atrial stimulation

• Symptomatic SSS 3

• Clinically relevant sinus bradycardia limiting BB therapy 4

Clinical need for AV-sequential pacing

• AV-block indicating AV sequential bradypacing 5 OR Atrial monitoring

• Patients without current need for oral anticoagulation, but ≥ 65 y of age, or those with elevated risk of stroke 6

CRT-DX considered 7,8,9

Yes

CRT-DX considered

• No need for atrial pacing 7

• Permanent AF but SR may be expected 8(i.e. planned cardioversion or spontaneous conversion may occur)

• Upgrade from DX ICD 9 No Indication for atrial

stimulation 3,4

Fig. 3. Proposed flowchart of device selection for ICD recipients, focusing on the DX technology. References used in the flowchart: 1. [41 , 43] ; 2. [42 , 43 , 44] ; 3. [26 , 41 , 42] ; 4. [41 , 43] ; 5. [41 , 42] ; 6. [22 , 29 , 31-33] ; 7. [1-5 , 9-12 , 16-18] ; 8. [21] ; 9. [22] ; 10. [1-5 , 8-14 , 23-28 , 30 , 34-37] ; 11. [15 , 19 , 20 , 23-26] . Abbreviations used in the flowchart: AF: atrial fibrillation, AV: atrioventricular, BB: beta-blocker, CRT: cardiac resynchronization therapy, ICD: implantable cardioverter defibrillator, LBBB: left bundle branch block, LVEF:

left ventricular ejection fraction, SR: sinus rhythm, SSS: sick sinus syndrome.

§ Is thereacurrentorexpectedneedforatrialstimulationre- quiringaseparateatriallead?

• Clinicallyrelevantsicksinussyndrome(SSS)orchronotropicin- competencerequiringatrialpacing

• Clinicallyrelevantsinusbradycardialimitingbeta-blockerther- apyinitiationoruptitration

§ Could the patient benefit from an additional atrial sensing dipole(DX)?

• Therapeuticconsequence ofthedetection ofa silent atrialar- rhythmia

• NeedforAVsequentialrightventricularpacemakerstimulation

• Permanent AF butsinus rhythmmay return(i.e. plannedcar- dioversionorspontaneousconversionmayoccur)

§ IstheatrialsensingoftheformerlyimplantedDXICDsystem stableenoughincaseoftheneedforanupgradetoCRT-D?

TohelptheselectionoftheICD typetobeimplanted,wehave developed adecision-makingflowchart, whichisshowninFig. 3. ThedecisionflowchartdefinesthepotentialindicationsofDXtech- nology.Inthesecases,besidestheDXtechnology,thefullscopeof algorithms,functions,physicalproperties,andeconomicaspectsof theavailableICDsystemshavetobeevaluatedandthedevicewith thehighestpotentialbenefitforthepatienthastobeimplanted.

Conclusion

Due toreliableatrialsensing, theDXICDsystemoffers anad- ditionalatrialintracardiacelectrogram,withtheearlydetectionof atrial arrhythmias,possiblyimprovedsupraventriculartachycardia discriminationandatrioventricularsequentialpacinginsingle-lead devices. Moreover, the CRT-DXsystem provides the possibility of cardiac resynchronizationtherapy via2leads. DXtechnologymay lead to lower procedural complexity,a lower radiation dose, and

lowerimplantcomplicationsduetothelackofanadditionalatrial lead.Basedontheavailable scientificevidence,theuseoftheDX ICD andCRT-DXsystemsmaybe a reasonableoptionincarefully selectedpatients,butfurtherclinicalresearchiswarrantedregard- ingthisnewtechnology.

DeclarationofCompetingInterest

M. V. reports consulting/lecture fees and/or nonfinancialsup- port from Abbott, Bayer, Biotronik, Daiichi Sankyo, Egis, Minimal lnvasiveTechnologyLtd.,andSanofi-Aventis.M.N.reportsconsult- ing/lecturefeesand/ornonfinancialsupportfromBiotronik,Boston Scientific, Egis, MicroPort, and Medtronic. T.B. is an employee of Biotronik Hungary Ltd. E. Z.reports consultancy/speaker/advisory board fees and/or nonfinancial support from Abbott, Amomed, Bayer,Biotronik,BostonScientific,Innomed, Orion,Pfizer, Replant, andRichter. G. Z.D. reportsconsulting/lecture feesand/or nonfi- nancialsupportfromAbbott,Biotronik,DaiichiSankyo,Medtronic, MSD,Replant,Richter,Pfizer,andSanofi.

Funding

Thisresearchdidnotreceiveanyspecificgrantorfinancialsup- portfromBiotronik.

References

[1] Chen BW , Liu Q , Wang X , Dang AM . Are dual-chamber implantable cardiovert- er-defibrillators really better than single-chamber ones? A systematic review and meta-analysis. J Interv Card Electrophysiol 2014;39:273–80 .

[2] Vamos M , Erath JW , Benz AP , Bari Z , Duray GZ , Hohnloser SH . Incidence of cardiac perforation with conventional and with leadless pacemaker sys- tems: a systematic review and meta-analysis. J Cardiovasc Electrophysiol 2017;28:336–46 .

[3] Perez AA , Woo FW , Tsang DC , Carrillo RG . Transvenous lead extrac- tions: current approaches and future trends. Arrhythm Electrophysiol Rev 2018;7:210–17 .

(6)

[4] Polyzos KA , Konstantelias AA , Falagas ME . Risk factors for cardiac implantable electronic device infection: a systematic review and meta-analysis. Europace 2015;17:767–77 .

[5] Duray GZ , Schmitt J , Cicek-Hartvig S , Hohnloser SH , Israel CW . Complications leading to surgical revision in implantable cardioverter defibrillator patients:

comparison of patients with single-chamber, dual-chamber, and biventricular devices. Europace 2009;11:297–302 .

[6] van Dijk VF , Boersma LV . The subcutaneous implantable cardioverter defibril- lator in 2019 and beyond. Trends Cardiovasc Med 2020;30:378–84 .

[7] Priori SG , Blomström-Lundqvist C , Mazzanti A , Blom N , Borggrefe M , Camm J , et al. ESC Guidelines for the management of patients with ventricular ar- rhythmias and the prevention of sudden cardiac death: the Task Force for the Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death of the European Society of Cardiology (ESC) Endorsed by: association for European Paediatric and Congenital Cardiology (AEPC). Eu- ropace 2015;17:1601–87 .

[8] Mittal S . The VDD ICD lead: friend or foe? Indian Pacing Electrophysiol J 2020;20:135–6 .

[9] Worden NE , Alqasrawi M , Krothapalli SM , Mazur A . Two for the Price of One":

a single-lead implantable cardioverter-defibrillator system with a floating atrial dipole. J Atr Fibrillation 2016;8:1396 .

[10] Dewland TA , Pellegrini CN , Wang Y , Marcus GM , Keung E , Varosy PD . Dual-chamber implantable cardioverter-defibrillator selection is associated with increased complication rates and mortality among patients enrolled in the NCDR implantable cardioverter-defibrillator registry. J Am Coll Cardiol 2011;58:1007–13 .

[11] Peterson PN , Varosy PD , Heidenreich PA , Wang Y , Ta Dewland , Curtis JP , et al. Association of single- vs dual-chamber ICDs with mortality, readmissions, and complications among patients receiving an ICD for primary prevention.

JAMA 2013;309:2025–34 .

[12] Ghani A , Delnoy PP , Ramdat Misier AR , Smit JJJ , Adiyaman A , Ottervanger JP , et al. Incidence of lead dislodgement, malfunction and perforation during the first year following device implantation. Neth Heart J 2014;22:286–91 . [13] Fontenla A , Salguero R , Martinez-Ferrer JB , Rodriguez A , Alzueta J , et al. Atrial

rate-responsive pacing and incidence of sustained atrial arrhythmias in pa- tients with implantable cardioverter defibrillators. Pacing Clin Electrophysiol 2016;39:548–56 .

[14] Biffi M , D’Onofrio A , Pignalberi C , Pisanò EC , Iacopino S , Curnis A , et al. Rate-responsive pacing and atrial high rate episodes in cardiac resyn- chronization therapy patients: is low heart rate the key? Clin Cardiol 2019;42:820–8 .

[15] Vamos M , Erath JW , Bari Z , Vagany D , Linzbach SP , Burmistrava T , et al. Effects of upgrade versus de novo cardiac resynchronization therapy on clinical re- sponse and long-term survival: results from a multicenter study. Circ Arrhythm Electrophysiol 2017;10:e004471 .

[16] Olshansky B , Richards M , Sharma A , Wold N , Jones P , Perschbacher D , et al. Survival after rate-responsive programming in patients with cardiac resynchronization therapy-defibrillator implants is associated with a novel pa- rameter: the heart rate score. Circ Arrhythm Electrophysiol 2016;9:e003806 . [17] Martin DO , Day JD , Lai PY , Murphy AL , Nayak HM , Villareal RP , et al. Atrial sup-

port pacing in heart failure: results from the multicenter PEGASUS CRT trial. J Cardiovasc Electrophysiol 2012;23:1317–25 .

[18] American Heart Association (AHA)EHRA/HRS expert consensus statement on cardiac resynchronization therapy in heart failure: implant and follow-up rec- ommendations and management. Europace 2012;14:1236–86 .

[19] Biffi M , Massaro G , Candelora A , Angeletti A , Valzania C , Martignani C , et al. Less is more: can we achieve cardiac resynchronization with 2 leads only? Int J Cardiol 2017;249:184–90 .

[20] Shaik NA , Drucker M , Pierce C , Duray GZ , Gillett S , Miller C , et al. Novel two-lead cardiac resynchronization therapy system provides equivalent CRT re- sponses with less complications than a conventional three-lead system: results from the QP ExCELs lead registry. J Cardiovasc Electrophysiol 2020;31:1784–92 . [21] Gasparini M , Steinberg JS , Arshad A , Regoli F , Galimberti A , Rosier A , et al. Re- sumption of sinus rhythm in patients with heart failure and permanent atrial fibrillation undergoing cardiac resynchronization therapy: a longitudinal obser- vational study. Eur Heart J 2010;31:976–83 .

[22] Biffi M , Iori M , De Maria E , Bolognesi MG , Placci A , Calvi V , et al. The role of atrial sensing for new-onset atrial arrhythmias diagnosis and management in single-chamber implantable cardioverter-defibrillator recipients: results from the THINGS registry. J Cardiovasc Electrophysiol 2020;31:846–53 .

[23] Gold MR , Ahmad S , Browne K , Berg KC , Thackeray L , Berger RD . Prospective comparison of discrimination algorithms to prevent inappropriate ICD therapy:

primary results of the Rhythm ID Going Head to Head Trial. Heart Rhythm 2012;9:370–7 .

[24] Friedman PA , Bradley D , Koestler C , Slusser J , Hodge D , Bailey K , et al. A prospective randomized trial of single- or dual-chamber implantable car- dioverter-defibrillators to minimize inappropriate shock risk in primary sud- den cardiac death prevention. Europace 2014;16:1460–8 .

[25] Kolb C , Sturmer M , Sick P , Reif S , Davy JM , Molon G , et al. Reduced risk for inappropriate implantable cardioverter-defibrillator shocks with dual-chamber therapy compared with single-chamber therapy: results of the randomized OP- TION study. JACC Heart Fail 2014;2:611–19 .

[26] Wilkoff BL , Fauchier L , Stiles MK , Morillo CA , Al-Khatib SM , Almendral J , et al. HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal im- plantable cardioverter-defibrillator programming and testing. Heart Rhythm 2016;13:e50–86 .

[27] Safak E , Schmitz D , Konorza T , Wende C , De Ros JO , Schirdewan A , et al. Clin- ical efficacy and safety of an implantable cardioverter-defibrillator lead with a floating atrial sensing dipole. Pacing Clin Electrophysiol 2013;36:952–62 . [28] Worden NE , Alqasrawi M , Mazur A . Long-Term Stability and Clinical Utility of

Amplified Atrial Electrograms in a Single-Lead ICD System with Floating Atrial Electrodes. Pacing Clin Electrophysiol 2016;39:1327–34 .

[29] Thomas G , Choi DY , Doppalapudi H , Richards M , Iwai S , Daoud EG , et al. Sub- clinical atrial fibrillation detection with a floating atrial sensing dipole in sin- gle lead implantable cardioverter-defibrillator systems: results of the SENSE trial. J Cardiovasc Electrophysiol 2019;30:1994–2001 .

[30] Kurt M , Jathanna N , Babady M , Schmidt J , Müller P , Gerguri S , et al. Avoiding inappropriate therapy of single-lead implantable cardioverter-defibrillator by using atrial-sensing electrodes. J Cardiovasc Electrophysiol 2018;29:1682–9 . [31] Hindricks G , Potpara T , Dagres N , Arbelo E , Bax JJ , Blomström-Lundqvist C ,

et al. ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association of Cardio-Thoracic Surgery (EACTS) [published online ahead of print, 2020 Aug 29]. Eur Heart J 2020:ehaa612 .

[32] Bertaglia E , Blank B , Blomström-Lundqvist C , Brandes A , Cabanelas N , Dan A , et al. Atrial high-rate episodes: prevalence, stroke risk, implications for man- agement, and clinical gaps in evidence. Europace 2019;21:1459–67 . [33] Biffi M , Ziacchi M , Ricci RP , Facchin D , Morani G , Landolina M , Lunati M , Ia-

copino S , Capucci A , Bianchi S , Infusino T , Botto GL , Padeletti L , Boriani G . Can we predict new AF occurrence in single-chamber ICD patients? Insights from an observational investigation. Int J Cardiol 2017;230:275–80 .

[34] Sticherling C , Zabel M , Spencker S , Meyerfeldt U , Eckardt L , Behrens S , et al. Comparison of a novel, single-lead atrial sensing system with a du- al-chamber implantable cardioverter-defibrillator system in patients without antibradycardia pacing indications: results of a randomized study. Circ Ar- rhythm Electrophysiol 2011;4:56–63 .

[35] Stazi F , Mampieri M , Cardinale M , Laudadio MT , Gargaro A , Battista del Giu- dice G . Implant and long-term evaluation of atrial signal amplification in a single-lead ICD. Pacing Clin Electrophysiol 2012;35:1119–25 .

[36] Safak E , D Ancona G , Kaplan H , Caglayan E , Kische S , Öner A , et al. New generation cardioverter-defibrillator lead with a floating atrial sensing dipole:

long-term performance. Pacing Clin Electrophysiol 2018;41:128–35 .

[37] Iori M , Giacopelli M , Quartieri F , Bottoni N , Manari A . Implantable cardioverter defibrillator system with floating atrial sensing dipole: a single-center experi- ence. Pacing Clin Electrophysiol 2014;37:1265–73 .

[38] Torbey E , Love CJ . Failure to shock: what is the mechanism? Pacing Clin Elec- trophysiol 2018;41:669–71 .

[39] Verma KP , Adam D . An unusual source of electromagnetic interference. Pacing Clin Electrophysiol 2018;41:1381–3 .

[40] Konstantino Y , Kleiman A , Amit G . Home monitoring report from a single lead Lumax DX implantable cardioverter defibrillator: new observations in a new system. J Arrhythm 2016;32:218–19 .

[41] Brignole M , Auricchio A , Baron-Esquivias G , Bordachar P , Boriani G , Brei- thardt OA , et al. ESC Guidelines on cardiac pacing and cardiac resynchro- nization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collab- oration with the European Heart Rhythm Association (EHRA). Eur Heart J 2013;34:2281–329 .

[42] Kusumoto FM , Schoenfeld MH , Barrett C , Edgerton JR , Ellenbogen KA , Gold MR , et al. ACC/AHA/HRS Guideline on the Evaluation and Management of Patients With Bradycardia and Cardiac Conduction Delay: a Report of the American Col- lege of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation 2019;140:e382–482 . [43] Ponikowski P , Voors AA , Anker SD , Bueno H , Cleland JGF , Coats AJS , et al. ESC

Guidelines for the diagnosis and treatment of acute and chronic heart fail- ure: the Task Force for the diagnosis and treatment of acute and chronic heart failure of the European Society of Cardiology (ESC)Developed with the spe- cial contribution of the Heart Failure Association (HFA) of the ESC. Eur Heart J 2016;37:2129–200 .

[44] Khurshid S , Epstein AE , Verdino RJ , Lin D , Goldberg LR , Marchlinski FE , Frankel DS . Incidence and predictors of right ventricular pacing-induced car- diomyopathy. Heart Rhythm 2014;11:1619–25 .

Ábra

Fig.  1. (Graphical  abstract)  Overview of  the  different  DX ICD systems and their  potential benefits  in different clinical situations
Fig. 3. Proposed flowchart of device selection for ICD recipients, focusing on the DX technology

Hivatkozások

KAPCSOLÓDÓ DOKUMENTUMOK

A comparison of risk stratification schemes for stroke in 79,884 atrial fibrillation patients in general practice. Atrial fibrillation as an independent risk factor for stroke: the

Heterogeneous photocatalysis showed the highest efficiency regarding the transformation of intermediates in presence of dissolved oxygen, while thiacloprid transformed with the

c Semmelweis University, Department of Oral and Maxillofacial Surgery, Maria u. The axial load was controlled at 2.0 kg. Bovine ribs were used as test models. In the guided group

severely impaired left atrial appendage function (which increases the risk of embolic stroke) is not common in patients with PAF. D) Peak left atrial appendage emptying flow

However, their well-known limitations, which include interactions with numerous foods and drugs as well as the need for frequent coagulation monitor- ing and dose adjustments,

The addition of 1% CNTs to the 316L steel yielded an increase in the hardness which can be explained by the smaller grain size, the higher dislocation den- sity and the

In addition to Raman measurements the high resolution transmission electron microscopy (HRTEM) studies have shown that the structure of MWCNTs remains intact after milling.. ©

The aim of second experiment was to determine the changes in endothelin-1, in N-terminal fragment of atrial natriuretic peptide (NT-ANP) and in atrial