• Nem Talált Eredményt

IMPEDANCE METER FOR SEMICONDUCTOR SURFACE CAPACITY MEASUREMENTS

N/A
N/A
Protected

Academic year: 2022

Ossza meg "IMPEDANCE METER FOR SEMICONDUCTOR SURFACE CAPACITY MEASUREMENTS "

Copied!
4
0
0

Teljes szövegt

(1)

IMPEDANCE METER FOR SEMICONDUCTOR SURFACE CAPACITY MEASUREMENTS

By

J. M. S06s and M. FULOP

Department of Physical Chemistry. Technical University Bndapest Received April 16, 1971

Presented by Prof. Dr. Gy. Varsanyi

Introduction

A useful method of investigating chemical changes, such as chemi- sorption, at semiconductor-electrolyte interfaces is to measure the space- charge capacity. The space-charge capacity of a semiconductor is a function of the surface-charge capacity, ·which is subject to changes in thc charge transition processes taking place on the surface.

Space-charge capacity is measured at frequencies above 104 Hz [1,

21,

because the part of the ci!·cuit at the interface can then be thought of as being replaced by a simple electric unit consistin g of a condensor and a resistor connected in series, which makes the space-charge capacity easy to determine.

In many interface studies it is also necessary to obtain a measure of the frequency-dependence of the space-charge capacity. It is an important re- quirement for any instrument used for such purposes that the tested semicon- ductor electrode can be put earth. Besides for the sake of determining the capacity at the required accuracy impedances ·with phasf' angle less than 45°

should be measured precisely. Only signals oflow (several millivolts) amplitude can be used for the measurements, else the polarization conditions at the interface (determined by the polarization voltage applied during the measure- ment) will be considerably affected. Lastly, in order to make the time de- pendence of thc capacity under given conditions accessible to study it to carry out many measurements in a short time, the measurement process itself must hc fast and easy to perform.

As the common measuring deviccs do not meet all the abovc requirements we have designed and built a "Z-meter" free from this deficiency. This paper reports on the meter's principle of operation and its accuracy. The measm:ing conditions and our own studies on phenomena at semiconductor-electrolyte interfaces will be discussed in subsequent publications.

(2)

136 J . .1£. sous and M. FOLO],

Principle of operation

The instrument works as follows (see Fig. 1): Impedance Zm to he measured is connected to the output terminals of generator G across the variahle standard resistance RH and dropping resistors Ro- C and Rp as well as the two identical resistors Rare hoth connected in series in the same manner.

A phase indicator is connected to points AB or BC. A vector diagram of the signals ohtained is shown in Fig. 2. The capacitance of C is chosen so that its impedance is equal to Rp at the measuring frequency, i.e. the voltage vector

0

C R

C Urn

Zen F R

E

Fig. 1 D

A~ ______ ~ ____ ~~B

E

Fig. 2

U CB is exactly perpendicular to measuring signal U DE. If RH is adjusted so that the phase angle hetween UAC and UCB is zero, URH and Uz will he equal, hence Z

=

RH. If the input terminals of the phase indicator are con- nected hetween AE and CF, the position of the slider on Rp can he adjusted so that U AE and U CF will he parallel, hence their phase angle wiIllikewise he zero.

The potentiometer scale can therefore he directly calihrated in q;-values giving a direct reading of the phase angle.

Construction of the apparatus

The phase indicator is an oscilloscope tuhe; the signals to the horizontal and vertical deflecting plates are provided hy selective amplifiers with dif- ferential inputs. These inputs are connected to points AB and CB when

(3)

~E.\lICOSDUCl'OR SURFACE CAPACITY 137

measuring Z and to points AE and CF when measuring rp. The gain of the selective amplifiers is high enough to permit the fullmodulatioIl of the electron Leam OIl the screen even with an input signal of 1m V. An ellipse is seen on the screen if the circuits are unbalanced on either side. The equilibrium is found by varying RH or Rp until the cllipse is closed to a 4.5° straight line.

As can be seen from the vector diagram, balancing bcgins with the deter- mination of Z. There is, however, no need to correct RH during the measure- ment, eliminating the lengthy iterated process of minimization proper to con- ventional measuring bridges. Moreover the iteration cannot become divergent and thus Z and rp are dcterminable, even for small phase angles. Another ad- vantage over the conventional Z-meters (Grutzmacher bridges) is that one of

0

C R

C H

R

E

Fig. 3

thc electrodes of thc measuring cell can be permancntly earthed, thus, shielding is casier and cven a 20 to 30 MHz measuring frcqucncy can Lc ohtaincd hy using a low-capacity RH and Rp.

To changc the measuring frequency, thc frequencies of thc measuring- signal gcncrator and both selective amplifiers as well as thc value of C must hc changed simultaneously. As our measuring gencrator covers the 160 Hz to 320 kHz frequency range in II steps, a great number of elements must run exactly together, since the phase transmission of both selective amplifiers must bc identical. Hence the following calihrating stages are involved in tuning thc ahove units: 1. All the selective amplifier inputs are connected to points CE, i.e. in each position the built-in tuning capacitors must be regulated until the ellipse on the oscilloscope screen closes, indicating identical phase transmission;

2. Xc (reactance of capacitor C) and Rp are balanced by connccting the inputs of one of the selective amplifiers across points BC. With a zero voltage at one input of the other amplifier, a signal shifted by 90° compared to the measuring signal is applied to the other one from a series-connected potential divider consisting of a capacitor Co and resistor T. The circuit and its vector diagram are shown in Figs 3 and 4, resp. When XCo is by two orders of magnitude greater than T, Ur is perpendicular to the measuring voltage vector U DE at an

(4)

.I . .If. ;lOC)" "lid M. FULijl'

error of a few tenths of a degree sufficiently accurate adjustmenL eau be madc at all frequencies.

H

G Llro o c

Fig. 4

The equipment also contains a high output resistance current generator supplying the polarization current for the electrodes, a voltmeter and a pH meter. These units are wired in the conventional wav. l\Iain technical speci- fication of the equipment:

Measuring frequency Measuring signal amplitude Range of imp('danc(' meaSUn'llH'nt Error of impedance mea8Urf'ment Range (If phasf' angle meaSUff'l1H'tJ t Error in phase angle measun'lll('nt Polarization current;;

Polarization voltage !l1p<!sun'IlH'llt between

Acknowledgemenl

160 Hz to 320 kHz 1 mY to 50 mYe}f

10 Q to 1 l\H2 max. . 2 u/o

o

to 90)

rnax. ~_!-2 ~ ;) 2 ,IIA to 2 iliA

2 Y: -,- 2 (; fJ'

The authors are grateful to Dr. Janos Giber for making po:-sible the cOllstruction of the instrument in the course of the complex semiconductor research program under his direction, and for helping with valuable ach-ice concerning- layout and measurement principles.

Summary

A non-iterative, compensated, impedance-phase angle meter with an oscilloscope phase indicator has been designed and built for measurements of space-charge capacity at semicon- ductor (germanium)-electrolyte boundaries. The instrument can operate oyer a wide measuring range (160 Hz to 320 kHz), Maximum amplitude of the measuring signal is 10 mV. :Measuring error for both Z and rp is max. 0 ,

References 1. MElIUUl'iG, R.: Philips Res. Repts. 19, 323 (1964)

2. GOBRECHT, H. et al.: Berichte del' Bunscnges. Phys. Chem. 67, 2, 14·2 (1963)

J

anos JYIihMy S065 }

1502 Budapest P. O. B. 91. Hungary Mihaly FtTLOP

Hivatkozások

KAPCSOLÓDÓ DOKUMENTUMOK

The rate equations for excess stored charge and for the photon number in semiconductor lasers can be represented by a nonlinear equivalent circuit.. This circuit is built

The values of impedances in every circuit are considered for lines whether self or mutual, fault impedance, ground return path impedance, ground wire, tower

where H is the magnetic field intensity, B is the flux density, Jt is the total current density, J s is the source current density, E is the electric field

Measurements and analysis of electromagnetic disturbances generated by a prototype of a navigation mobile robot has been carried out.. Results of the measurements for the model of

Impedance cardiogra- phy (electrical velocimetry) and transthoracic echocardiography for non-invasive cardiac output monitoring in pediatric intensive care patients: a

a) There is an idealized surface or surface phase separating two homogeneous bulk phases (see Figures 1 and 2). The bulk phases are in equilibrium with the surface or surface phase.

The selection of alkali metal working fluids for Rankine cycle space power systems has necessitated investigation of two-phase heat transfer and fluid flow characteristics.. The

Table 2 below shows the values of total transfer capacity on the Slovakian - Czech cross border connection with installed phase shift transformer on line V 404. In