• Nem Talált Eredményt

of Molecular Weight

N/A
N/A
Protected

Academic year: 2022

Ossza meg "of Molecular Weight"

Copied!
20
0
0

Teljes szövegt

(1)

CHAPTER 21

Microdetermination of Molecular Weight

The microdetermination of the molecular weight of a substance may be accom­

plished by several different methods,* most of which are small-scale applica­

tions of the various macromethods found throughout the literature, such as the ebullioscopic (elevation of boiling point of the solvent), cryoscopic (depres­

sion of the freezing or melting point of the solvent), and the vaporimetric (vaporization of the substance in a closed system) methods.4 1 T o a certain ex­

tent, the determination of the neutralization equivalent is one of molecular weight, as explained in Chapter 15. The same holds true of the saponification equivalent3 7 (see Chapter 2 0 ) and the hydrogénation number (see Chapter 1 9 ) . In a broad sense, the determination of any single element or group can be interpreted as that of the molecular weight. For example, if a substance is found to contain 1 0 % of nitrogen, its molecular weight must be 140 if only one nitrogen atom is present or 280 if two are present, etc. In other words, it is 140 or some multiple of this figure. Similarly, if a substance contains 1 0 % of acetyl, its molecular weight is 4 3 0 or some multiple.

Two methods of determining the molecular weight will be described in detail in this chapter, namely, the R a s t6 6'6 7 and the Signer7 5 as modified by Clark.1 5-3 6 Both methods have advantages and disadvantages and, in the opinion of the author, neither can be considered to be absolutely reliable. Therefore, it is advisable to use both methods and to use more than one type of solvent with each method, thereby reducing the possibility of drawing erroneous con­

clusions from the results. The possibility of decomposition, dissociation, reac­

tion, polymerization, etc., during the determination should not be ignored.

The Rast method is more frequently used because of the time factor. How­

ever, the results of collaborative study5 4 indicated that the Signer method is the better.

THE RAST METHOD FOR THE DETERMINATION OF MOLECULAR WEIGHT

The Rast m e t h o d ,1 6'2 9-3 0'5 0'5 1-6 6-6 7'7 0-7 3'7 9 a cryoscopic one, for the determina­

tion of molecular weight, is based upon the well-known principle that a solute depresses the melting or freezing point of a solvent proportional to the mol

528

* Please see references 4, 5, 16, 27, 29, 30, 50, 51, 63, 66, 67, 7 0- 7 3 , 79.

(2)

529 Rast Method

fraction of solute present.4 1 However, the solute or substance whose molecular weight is being determined, must not undergo decomposition at the tempera­

ture of the melting point of the solvent and also there must be no reaction between the two substances. The solvent selected should melt at some tempera­

ture below the melting or boiling point of the sample to be analyzed.

Reagents

Before any of the solvents listed below can be used for a determination with an unknown, they must be used with a test sample such as azobenzene or anthracene going through the regular procedure to obtain their constants (see below).

CAMPHOR, M.P., 176°-180° C .8 2

Very pure camphor is required as a solvent for high-melting samples. It should be resublimed in the laboratory and stored in a wide-mouthed ground glass- stoppered bottle.

EXALTONE (CYCLOPENTADECANONE)82 M.P., 65.6° C.

This material should be very pure. It is used as a solvent for the substances with melting points below that of camphor.

OTHER SOLVENTS-SEE TABLE 36

Camphene, m.p. 4 9 ° C. ; borneol, m.p. 206° C. ; bornylamine, m.p. 164° C. ; camphoquinone, m.p. 190° C. ; cyclohexanol, m.p. 2 4 . 6 ° - 2 4 . 8 ° C. ; cam- phenilone, m.p. 38° C ; and pinene dibromide, m.p. 170° C. may be used following the rule in regard to the relation between melting points of the solute and solvent given above. For additional solvents the reader is referred to the list of twenty-seven compounds compiled by R o t h7 1-7 2 that includes com­

ments on the use of each as well as references for their preparation and purification.

Apparatus

MELTING POINT APPARATUS37 8 2

The ordinary mechanically stirred melting point bath containing Crisco* or silicone fluid is satisfactory for the determination. It should be well illuminated.

A set of good grade of Anschiitz thermometers is required, one thermometer for each of the ranges in which fall the melting points of the solvents used.

A satisfactory reading lens is used for observing the thermometer. Figure 203

* Hydrogenated vegetable o i l .6 4

(3)

21. Molecular Weight 530

shows a new commercially available8 2 melting point apparatus particularly suitable for the determination, when the bath assembly modified for molecular weight determinations (Fig. 203b) is used. This bath assembly, which in­

cludes an immersion heater, stirrer, armored thermometer, and capillary tube

FIG. 203. Thomas-Hoover melting point apparatus, showing enlarged view of peri- scopic reading device, a, and special bath assembly for molecular weight determinations by Rast method, b.

holder, fits into a well in the top of the control cabinet. A periscopic ther­

mometer reader provides magnified thermometer indication in a window ap­

proximately 1.5 inches above the center of the capillary viewing lens so that both the temperature and the capillary contents may be under observation with little shifting of the operator's eyes. This adds to the accuracy of the method.

The bath is silicone fluid. It heats rapidly with no lagging, the heat transfer allowing the bath temperature to rise to 350° C. in 6.5 minutes. Rapid cooling

(4)

531 Rast Method

is affected by means of a compressed air line, making it possible to cool from 300° C. down to 150° C. in about 11 minutes.

MELTING POINT CAPILLARIES16'2d>30>50>51>66>67>70-7S>7d

Pyrex2 1 glass tubing is drawn out and sections with tapering walls should be selected having dimensions approximately those shown in Fig. 204, a. The

* 3 m m .

Ε Ε m

Ε Ε if) ω

Ε Ε

Ο

ο

Ε Ε

1 '

" S E A L E D

. S A M P L E + S O L V E N T

—Η 1^2 mm.

FIG. 204. Capillaries for microdetermination of molecular weight by the Rast method, showing various stages—details of construction, dimensions are approximate.

smaller ends are sealed off as shown. Slightly longer and narrower sections are also selected which will fit into the closed outer melting tubes, a, as shown in Fig. 204, b. Two open ends tubes, b, are required for each closed end tube, a, per sample. Both open ends of the longer tubes, b, are fire polished. A glass rod is drawn out and sections selected that are a few centimeters longer than the open tubes, b, but of small enough outside diameter to pass through them as shown in Fig. 204, c. These small glass rods act as plungers to force the samples from the open end tubes, b, into the melting tubes, a, as described

(5)

21. Molecular Weight 532

below under Procedure. A number of these units should be prepared—two closed tubes, a, three open end tubes, b, and two plunger rods being required for a molecular weight determination, not run in duplicate on solid substances.

When determining the molecular weight of liquids, the glass rod is not needed but the open end tube should be drawn out into a capillary at the smaller end.

Capillaries of slightly different dimensions are commercially available.8 2

Procedure

In the p r o c e d u r e1 6'2 9'3 0'5 0-5 1'6 6-6 7-7 0-7 3'7 9 the closed end capillary, a, is weighed on the balance. The solid sample is placed on a small watch glass and the smaller tip of the open end tube, b, is tapped several times into the sample so that the material packs tightly in it at this point. The excess sample is wiped from the outside of the open end tube, b. The open end tube, b, is inserted in the weighed closed end one, a, so that the sample end is near the closed bottom. The plunger rod is passed through the inner capillary so that the sample is forced into the bottom of the weighed tube, a. Both open end tube, b, and the plunger rod are removed together, care being exercised so that no particles of sample are carried along the upper wall of the closed tube. The tube containing the sample is returned to the balance and reweighed to obtain the weight of the sample. (Note: With liquid samples no plunger rod is used and the smaller tip of the open end tube is drawn out into a fine capillary to aid in the insertion of the sample.) A second open end tube, b, is used along with a second glass rod for insertion of the solvent on top of the sample in the closed tube, exactly in the same manner as for the intro­

duction of the sample. The amount of solvent depends on the molecular weight of the substance, but in general ten to twenty times the weight of the sample is used. The closed end tube, a, is returned to the balance and weighed.

The three weighings (empty tube ; tube -\- sample ; tube -)- sample -|- solvent) give the weights of sample and solvent. (For substances melting above the melting point of camphor, this solvent is used and exaltone is used for the others.

If, however, neither of these is a solvent for the sample, which must be determined previously, one of the other solvents listed may be used.) The weighed melting point tube is removed from the balance and sealed off by means of a needle-point flame from a blast lamp at a distance of about 25 mm.

from the bottom as shown in Fig. 204, d. ( T h e tube is allowed to collapse in the flame to form a solid portion which is then given a slight twist to ensure its being sealed.) The upper part of the tube above the seal is kept attached to the bottom for use as a handle. Great care must be exercised in the sealing off process so that neither sample nor solvent is lost by the heat­

ing or the determination is ruined.

(6)

533 Rast Method

A second sample capillary is filled as above with approximately the same amount of solvent, but no weighing is necessary since this tube contains no sample and is used only for the purpose of obtaining the melting point of the solvent.

The two sealed tubes (one with solvent alone, the other with solvent plus sample) are attached, side by side, to an Anschiitz thermometer of proper range in the manner usually prescribed for the determination of melting points3 7-5 6 (material in the sample tube at the same level as the bulb—about one-half way down in the bath). The bath is stirred rapidly and heated to whatever temperature is required to completely melt the mixture. It is cooled a few degrees to allow the melt to resolidify and then slowly reheated so that the temperature rises at the rate of about 0.5° C. per minute. The temperature at which the last crystal disappears is taken as the melting point of the mixture.

The temperature is increased at the same rate until the solvent has melted. This too is that point at which the last crystal disappears. The bath is cooled a little below the melting point of the mixture and when it has resolidified, the heating is repeated at the same rate to obtain the melting points of the mix­

ture and of the solvent. This process of heating, cooling, reheating, etc., is repeated until successive determinations check to within ± 0 . 2 ° C. The repeti­

tion is necessary since this is the only means of knowing when the mixture is homogeneous (solution is complete). Even though the one tube contains only solvent, the duplication of the melting point of this material is quite in order since it gives practice as well as a check on the value. ( I f the freezing point of the mixture is more clearly defined than the melting point, which is sometimes the case, the former may be noted instead of the latter. In such a case, the temperature is raised to a few degrees above the melting point and the bath cooled at the rate of 0.5° C. per minute. Obviously, the freezing point of the solvent must also be recorded rather than the melting point in such cases.) Since the melting (or freezing) points of both the solvent and the mixture are done simultaneously, under identical conditions, there is no need for calibration or stem corrections due to immersion. Only the depression of the melting (or freezing) point of the solvent enters into the calculation—

not the actual temperatures.

Calculation:

The molecular weight is calculated from the equation:

M W = 1000 X * X 5

L X At

where Κ = the molecular melting (or freezing) point depression. (Melting or freezing depression for one mol of solute per 1000 grams of solvent.)

S = weight of sample (solute) L = weight of solvent

A t = depression of the melting (or freezing) point of the solvent.

(7)

21. Molecular Weight 534

Table 36 gives the list of some possible solvents together with their melting points, molecular melting (or freezing) point depressions and a few com­

ments. The values for Κ should not be accepted as listed in the table, but should be determined by analysis of a standard test sample. (Since all other members of the equation are known in the test analysis, K, the only unknown, is calculated.) The standard test samples must be of the highest purity.

Acetanilide, azobenzene, chloroanthraquinone, anthracene, etc., may be used.

Examples:

a. The following data are obtained during the course of a determination using camphor (K = 39.1) as the solvent:

W t . sample = 1.004 mg.

Wt. camphor = 20.807 mg.

M.P. of camphor = 178.0° C.

M.P. of mixture = 166.2° C.

.*. At = Π 8 . 0 — 166.2 = 11.8° C.

. M W _ 1000 X 39.1 X 1.004 20.807 X 11.8

= 160

b. The following data are obtained from the same compound as above (used in Example a) but with a different lot of camphor (K = 38.2) :

Wt. sample = 2.073 mg.

W t . camphor = 38.632 mg.

M.P. of camphor = 177.5° C.

M.P. of mixture = 164.6° C.

.'. At = 12.9° C.

. 1000 X 3 8·2 X 2·0 7 3 . . M . W . =

38.632 χ 12.9

= 159

c. The following data are obtained from another compound using exaltone (K — 21.1) as the solvent:

W t . sample = 1.102 mg.

Wt. exaltone = 12.679 mg.

M.P. of exaltone = 65.5° C.

M.P. of mixture = 56.2° C.

/ . At = 9.3° C.

1000 X 21.1 X 1.102 M.W. =

12.679 X 9.3

= 197

The allowable error for the determination is ± 5 % of the molecular weight.

(8)

535 Signer Method

TABLE 3 6 * SOLVENTS

Solvent M . P . ( ° C . ) iC(°C.) C o m m e n t s1 6'7 1'7 2

B o r n e o l5 0'5 1'5 8-7 1'7 2 206 35.8

Bornylamineie.aMO.eQ, 7 0 - 7 3 164 40.6 Suitable for basic sub­

stances C a m p h e n e2 9'3 0-5 8'7 0-7 3 49 31.0

C a m p h e n i l o n e2 9'3° .6 2' ™ -7 3 38 64.0

Camphoquinoneief29,30,eo,ei,70-73 190 45.7 Used for high-melting substances

C a m p h o r t1 6'2 9'3 0-5 0'5 1'7 0-7 3 1 7 6 - 1 8 0 37.0-40.0 Used for high-melting substances

Cyclohexanol1 6'8 5 24.6-24.8 37.7 Unsuited for heavily

halogenated com­

pounds Exaltone ( Cyclopentadec-

a n o n e ) !1 6'2 9'3 0»7 0-7 3

65.6 21.3 Very good. Used for azo dyes, sterols, carote- noids, quinones

Pinene dibromide7 1'7 2 170 80.9

* For additional information, refer to the table of twenty-seven compounds compiled by Roth.7i.72

f Commercially available.8 2

THE SIGNER {ISOTHERMAL DISTILLATION) METHOD FOR THE DETERMINATION OF MOLECULAR WEIGHT

The Signer m e t h o d1 5*1 6'7 5'7 9 for the determination of molecular weight is based upon the principle of isothermal distillation. I f two solutions, having a common solvent, are placed in an evacuated system and held at a constant temperature, the solvent will distill from the solution of greater vapor pres­

sure to the one of lesser, until equilibrium is established. Then the volumes of the two solutions will be constant and equimolar. I f the one solution con­

tains a known substance and the other solution contains an unknown, the molecular weight of the latter can be calculated from the data obtained in such an experiment. The main disadvantage of the method is the time re­

quired for equilibrium to be established so that the calculations can be made.

After the apparatus has been filled and sealed off, however, a daily reading of the two volumes is all that is required.

(9)

2 1 . Molecular Weight 536

Reagents

SOLVENTS

Any pure volatile solvent with a high vapor pressure such as those listed below may be used, provided that both the unknown sample and the standard are sufficiently soluble in it to give solutions of slightly less than 0 . 1 molar concentration.

Obviously, there must be no reaction between the solvent and either (or both) the sample or the standard. Water is a poor solvent for these determinations.

STANDARDS

A variety of standards of highest purity may be used. Colored substances, as for example azobenzene, are preferred, since with them there is little or no chance of confusing the solution of the standard with that of the unknown (or vice versa) and labeling is unnecessary. Sometimes it is desirable to use as a standard a substance of similar structure to that of the unknown.

The apparatus7 4 used is that described by Clark,1 5-1 6 a diagram of which is shown in Fig. 2 0 5 , A. It consists of two graduated tubes, each of which is connected to a small bulb. The bulbs in turn are joined by an inverted U- shaped section, at the upper sides of which are two tubes used for introducing the sample, standard, and solvent as well as for evacuation of the system.

The capacity of the graduated tubes is 2 ml. each in the apparatus described by Clark. However, the author of this book has found those of 1-ml. capacity each to be quite useful when very small samples are used. With the smaller graduated tubes, the other dimensions are obviously reduced somewhat, with the exception of the sample inlet tubes. I f the latter are reduced in diameter, they are too small to accommodate the weighing tubes (Figs. 4 7 - 4 9 , Chap­

ter 3 ) .

VACUUM SYSTEM

A suitable vacuum system composed of a good electrically driven pump, traps, and manometer is required for evacuating the apparatus.

Ethyl ether Ethyl bromide Methylene chloride Ethyl formate

Acetone Methyl acetate Glacial acetic acid

Apparatus

(10)

537 Signer Method

BLAST LAMP

A blast lamp of the type used for sealing Carius combustion tubes (see Chap­

ter 1 0 ) is required for sealing off the inlet tubes.

Δ

FIG. 205. Signer (isothermal distillation) apparatus for determination of molecular weight. (A) Details of construction. (B, C, D) Various stages in filling procedure.

Procedure16 d

The sample is weighed in a weighing tube (Figs. 4 7 - 4 9 , Chapter 3 ) and introduced into the one bulb. The standard is weighed similarly and introduced into the other bulb. The relative amounts of sample and standard should be in proportion to their molecular weights, if that of the unknown can be esti­

mated. As stated above, the amounts of material should be such that solutions of slightly less than 0.1 molar in concentration will be obtained ( i f the sub­

stance is soluble enough—otherwise less material is used). Care should be exercised so that no particles remain on the walls of the inlet tubes. After introduction of the sample and the standard, both inlet tubes are constricted

(11)

21. Molecular Weight 538

down to capillaries of about 1 mm. I.D., through the use of a blast lamp flame in a manner similar to that employed for sealing the Carius combustion tubes (Chapter 1 0 ) . The constrictions are made close to the bottom of the inlet tubes near to the upper sides of the inverted U (Fig. 205, B). After cooling, solvent is added through each capillary into the bulbs containing the sample and the standard. When 2-ml. graduated tubes are employed slightly less than 2 ml. of solvent is added to each bulb, and, similarly, when the one- mi. size is used a little less than one ml. is added to each. The contents of the bulbs are chilled somewhat to reduce the danger of fire, and, after warm­

ing one of the capillaries gently, to remove adhering solvent, it is sealed off

ι ι ι ι ι ι ι ι ι ι ι ι ι ι ι ι ι ι

— η—σ—θ θ θ e

,,το I ι ι ι ι ι ι ι ι

ι

ι ι ι I

0 2 4 6 β 10 12 14 16

DAYS

FIG. 206. Curves showing time-volume relationships during determination of molecular weight of α-nitronaphthalene using azobenzene as standard. Solvent, acetone; solutions, 0.0932 molar at equilibrium. Molecular weight of α-nitronaphthalene: Calculated, 1 7 3 . 1 ; found, 173.7.

with a needle-point flame from a blast lamp (Fig. 205, C ) . The other inlet tube is attached to the vacuum system and the apparatus pumped out. As the apparatus is evacuated, solvent distills out which is accompanied by a lower­

ing of the temperature so that further external cooling is unnecessary. After the pressure in the system no longer drops, the apparatus is tilted so that the solutions flow into the graduated tubes. The contents of each tube is allowed to evaporate at the minimum pressure obtainable until each graduated tube is about one-half to two-thirds full (about 1-1.3 ml. with the 2-ml. tube and 0.5-0.65 ml. with the 1-ml. size). The actual reduction in volumes should occur during the time the solutions are in the bulbs and the unit should be tilted from time to time to note the volumes. The second capillary inlet tube is sealed by carefully applying the flame of the blast lamp at the constricted

(12)

539 Signer Method

point and drawing the upper portion away as soon as it is soft enough to do so (see Fig. 205, D ) . I f the evacuated capillary is heated too soft, there is danger of a hole being sucked in at this point, ruining the experiment. (It is sug­

gested that the beginner practice sealing off an evacuated capillary before per­

forming this part of the determination.)

The volumes in each graduated tube are recorded and the apparatus is set aside under isothermal conditions. I f the laboratory is adequately air-condi­

tioned day and night, it suffices to place the evacuated tubes in a closed container such as an aluminum pot or pressure cooker. Otherwise, the apparatus must be placed in a constant-temperature cabinet. ( I f a constant-temperature bath is used, the tubes must be dried quickly upon removal so that there is no chilling effect from evaporation of bath liquid on their surface.)

Each day thereafter the volumes of both graduated tubes are recorded.

If desired, a graph may be made plotting the volume of each against the time as shown in Fig. 2 0 6 . As time goes on, solvent will distill from the solution of greater vapor pressure to the one of lesser, as explained above. Finally, the volumes become constant and equimolar, and the experiment is completed.

Calculation:

The molecular weight of an unknown is calculated from the final data obtained by substitution in the following f o r m u l a1 6-7 9:

G.MV ΜΛ -

1 GV1

where

M = M . W . of standard

V - volume of solution of standard G = wt. of standard

M, = M.W. of unknown

Vi = volume of solution of unknown G, = wt. of unknown

Example:

10.876 mg. of an unknown and 13.013 mg. of azobenzene are used. At equilibrium (after 7 days) the volume of the solution containing the unknown is 1.65 ml. and that containing the azobenzene is 1.05 ml. (Note: M . W . of azobenzene = 182.228.)

10.876 X 182.228 X 1.05

1 13.013 X 1.65

= 97.

The allowable error for the determination is dz 5 % of the molecular weight.

(13)

21. Molecular Weight 540

ADDITIONAL INFORMATION FOR CHAPTER 21

It was stated in the beginning of this chapter that, in the opinion of the author, neither of the two methods described in detail is completely satisfactory.

Recently, the use of thermistors is attracting attention and one method has been described using them, which, from reports, appears quite interesting from both the standpoints of speed and accuracy.2 6-3 4'5 5'8 3* As this book goes to press, the unit in the author's laboratory has been in use only a few months, but the indications are that this method and apparatus will prove to be reliable for a variety of compounds.

This method described by Neumayer4 8 employs isothermal distillation and requires only 3 minutes for equilibrium. Pure solvent and solution are exposed to the vapor of the solvent. Two thermistors are used, one for the solvent and the other for the solution. The solvent temperature remains constant, being subject to equal rates of evaporation and condensation. The solution, however, increases in temperature due to vapor condensation. The temperature differ­

ence is proportional to the mol fraction of the solute. This method is described below.

Apparatus48 82

The Wheatstone bridge circuit used is shown in Fig. 207. The various com­

ponents of the circuit are as follows: BL9 4.5 volt battery; single-pole, single- throw switch; R± and R2, 100,000 ohm wire-wound fixed resistors; RS, 1,000 ohm wire-wound variable resistor; R±, 20 ohm wire-wound variable resistor;

RR„ 1,000 ohm ten-turn, wire-wound micropotentiometer ; TM1 and TM2, type G A5 1P 8 thermistors made by Fenwal Electronics, Inc., Framingham, Massachu­

setts; MLF 0 - 5 0 microammeter ; M2, Null Detector Model No. 1 0 4 W I G made by the Minneapolis-Honeywell Regulator Co., Minneapolis, Minnesota.

The solvent-vapor chamber and thermistor assembly are shown in Fig. 208.

The various parts are as follows: A, glass solvent-vapor chamber; B, Bakélite cap machined to fit Ί1 7 1 / 2 5 glass joint; C, 6-mm. stainless steel rod support for thermistor assembly; D, Teflon thermistor support; E, 5-mm. glass tubing;

F, Fenwal type G A5 1P 8 thermistors; G, cupric oxide-phosphoric acid cement;

H, 40-turn platinum wire coils made with 0.3-mm. diameter wire; /, 3-mm.

thick absorbent paper lining the total depth and about three-fourths of the circumference of the inner chamber wall and covered on both sides with 1.5- mm. mesh aluminum screen; ] , stainless steel crucible; K, solvent; L, constant temperature water bath.

The apparatus is now commercially available.8 2 This model has replaced the crucible with a small funnel through which the excess solution may be removed by suction. The medicine droppers are inserted in the cap of the chamber and kept there at all times to prevent temperature variations.

(14)

541 Additional Information for Chapter 21

FIG. 207. Neumayer apparatus for determination of molecular weight—Wheatstone bridge circuit.

FIG. 208. Neumayer apparatus for determination of molecular weight—solvent-vapor chamber vs. thermistor assembly.

P r o c e d u r e

The apparatus shown in Fig. 208 is immersed in a water bath thermostated at about 30° C. ±: 0.01. Solvent is poured into the vapor chamber so that there is about one cm. of excess solvent in the bottom of the chamber over that re­

quired to saturate the absorbent paper lining the inside wall of the chamber.

Both platinum wire coils covering the thermistors are rinsed with about 0. Ι ­ Ο. 2 ml. of solvent. The solvent can be readily placed on the coils by means of

(15)

21. Molecular Weight 542

an elongated medicine dropper inserted into the vapor chamber through a hole in the Bakélite cap. The medicine dropper is stored, when not in use, in copper tubing which is sealed on the bottom and immersed in the water bath. The micropotentiometer, R5, shown in Fig. 2 0 7 is set at zero resistance and switch, Sly is turned on. When temperature equilibrium has been estab­

lished within the vapor chamber, the bridge is balanced using resistors, JR3 and R±. The solvent in contact with the sensing thermistor, TMlt is replaced by rinsing with 0 . 1 - 0 . 2 ml. of solution which has been previously brought to 3 0 ° C . by immersion of its container in the water bath. The bridge is bal­

anced using the 1 0 0 0 ohm micropotentiometer.

Approximately 3 minutes are required for the sensing thermistor to reach its maximum temperature. The maximum resistance of the micropotentiometer is recorded and this value is used to calculate the molecular weight of the solute. The sensing thermistor is rinsed with solvent and the micropotentiometer is returned to zero resistance. Temperature equilibrium will again be estab­

lished in about 3 minutes, at which time the molecular weight of the next sample may be determined. I f desired, the sample may be recovered from the stainless steel crucible within the solvent-vapor chamber.

Calculation :48

Thermistor resistance varies logarithmically with temperature but only a maximum of approximately 1 % of the total resistance of one of the thermistors

1000

/

(OHMS)

S 0 0

0

0 aoo5 0.010 0.015

MOL F&ACTION

FIG. 209. Relationship of AR vs. mol fraction of solute (Neumayer method for determination of molecular weight).

(16)

543 Table of References

is utilized in these experiments so Δ Γ is essentially proportional to ΔΚ. It is also true that AR = Κ χ MF (see Fig. 2 0 9 ) , where AT and AR are the changes in temperature and resistance, respectively, of the sensing thermistors, Κ is a proportionality constant and MF is the mol fraction of solute in solu­

t i o n . *4 1 Substituting the appropriate terms for MF and rearranging yields the equation : f

g solute X MW solvent (K—AR) MW solute

AR X g solvent

where MW is molecular weight and g is weight in grams. Κ is determined for each solvent by making measurements on known amounts of test compounds.

T A B L E 37

ADDITIONAL INFORMATION ON REFERENCES:!: RELATED TO C H A P T E R 21 In addition to the above, as well as the procedures presented in detail in the preced­

ing pages of this chapter, the author wishes to call to the attention of the reader the references listed in Table 37. (See statement at top of Table 4 of Chapter 1, regarding completeness of this material.)

Books

Belcher and Godbert, 4, 5 Bonnar, Dimbat, and Stross, 7 Clark, E . P., 16

Clark, S. J . , 17 Furman, 27 Grant, 29, 30 Millard, 41

Milton and Waters, 42 Niederl and Niederl, 50, 51 Niederl and Sozzi, 52 Pregl, 63

Roth, 7 0 - 7 3 Steyermark, 79

Sucharda and Bobranski, 8 0 Collaborative study

Ogg, 54, 55 Review

Wilson, C. L., 84

* Mol fraction =

Ultramicro-, submicro-methods Brandstàtter, 9

Gibson and Cur rie, 28 Guerrant, 31

Kofler and Brandstàtter, 38

Ebullioscopic (boiling point rise) Belcher and Godbert, 4, 5

Colson, 19, 20 Dimbat and Stross, 24 Furman, 27

Grant, 29, 30 Magee and Wilson, 4 0 Niederl and Niederl, 50, 51 Pregl, 63

Rieche, 68, 69 Roth, 7 0 - 7 3

Sucharda and Bobranski, 80, 81 Wilson, 84

mois solute mois solute + mois solvent

f The slope of the line obtained by plotting the resistance against the concentration also may be used in calculating the results.8 3 a

t The numbers which appear after each entry in this table refer to the literature citations in the reference list at the end of the chapter.

(17)

2 1 . Molecular Weight 544

T A B L E 37 {Continued) Cryoscopic (depression of melting or

freezing point) Aluise, 1

Brandstâtter, 9 Csokan, 22 Furman, 27

Gibson and Currie, 28 Grant, 29, 30 Iwamoto, 36

Kofler and Brandstâtter, 38 Kubota and Yamane, 39 Niederl and Niederl, 50, 51 Pirsch, 57

Rast, 66, 67 Roth, 7 0 - 7 3 Smit, 76

Smit, Ruyter, and Van Wijk, 77 Steyermark, 79

Zscheile and White, 87 Using Kofler micro hot stage

Brandstâtter, 9 Gibson and Currie, 28 Kofler and Brandstâtter, 38 Osmotic pressure methods Christiansen and Jensen, 14 Forster and Breitenbach, 25 Effusion method

Nash, 45

Vaporimetric, vapor pressure compari­

son methods Blank, 6

Bratton and Loch te, 10 Furman, 27

Gysel and Hamberger, 32 Hawkins and Arthur, 33 Nash, 44

Nernst, 47

Niederl and Niederl, 50, 51 Puddington, 65

Isothermal distillation methods Barger, 2, 3

Bourdillon, 8 Childs, 13

Isothermal distillation methods (Conf.) Clark, E. P., 15, 16

Francis, 26 Guerrant, 31 Hofstaeder, 34 Hoyer, 35

Morton, Campbell, and Ma, 43 Neumayer, 48

Niederl and Levy, 49 Niederl and Niederl, 50, 51

van Nieuwenburg and van Ligten, 53 Ogg, 54, 55

Roth, 72 Spies, 78

White and Morris, 83 Wilson and Bini, 83a Wright, 86

Potentiometric, nonaqueous titration Brockmann and Meyer, 12

Neutralization equivalent See Chapter 15

Spectrophotometric method Cunningham, Dawson, and Spring, 23 Apparatus

British Standards Institution, 11 Christiansen and Jensen, 14 Clark, E. P., 16

Clark, S. J . , 17 Colson, 1 8 - 2 0 Dimbat and Stross, 24 Hawkins and Arthur, 33 Hoyer, 35

Nash, 4 5 , 46

Niederl and Niederl, 50, 51 van Nieuwenburg and van Ligten, 53 Roth, 7 0 - 7 3

Steyermark, 79 White and Morris, 83 Use of thermistors

Dimbat and Stross, 24 Francis, 26

Hofstaeder, 34 Neumayer, 48 Ogg, 55

Wilson and Bini, 83a

(18)

545 References

REFERENCES

1. Aluise, V . Α., Ind. Eng. Chem., Anal. Ed., 13, 365 ( 1 9 4 1 ) . 2. Barger, G , Ber., 37, 1754 ( 1 9 0 4 ) .

3. Barger, G., / . Chem. Soc, 85, 286 ( 1 9 0 4 ) .

4. Belcher, R., and Godbert, A. L., "Semi-Micro Quantitative Organic Analysis," Long­

mans, Green, London, New York, and Toronto, 1945.

5. Belcher, R., and Godbert, A. L., "Semi-Micro Quantitative Organic Analysis," 2nd ed., Longmans, Green, London, 1954.

6. Blank, E. W., Mikrochemie, 13, 149 ( 1 9 3 3 ) .

7. Bonnar, R. U., Dimbat, M., and Stross, F. H., "Number Average Molecular Weight,"

Interscience, New York, 1958.

8. Bourdillon, J . , / . Biol. Chem., 127, 617 ( 1 9 3 9 ) .

9. Brandstàtter, M., Mikrochemie ver. Mikrochim. Acta, 36/37, 291 ( 1 9 5 1 ) . 10. Bratton, A. C , and Lochte, H. L., Ind. Eng. Chem., Anal. Ed., 4, 365 ( 1 9 3 2 ) . 11. British Standards Institution, Brit. Standards, 1428, Pt. K l ( 1 9 5 8 ) .

12. Brockmann, H., and Meyer, E., Ber., 86, 1514 ( 1 9 5 3 ) . 13. Childs, C. E., Anal. Chem., 26, 1963 ( 1 9 5 4 ) .

14. Christiansen, J . Α., and Jensen, C. E., Acta Chem. Scand., 5, 849 ( 1 9 5 1 ) . 15. Clark, E. P., Ind. Eng. Chem., Anal. Ed., 13, 820 ( 1 9 4 1 ) .

16. Clark, E. P., "Semimicro Quantitative Organic Analysis," Academic Press, New York, 1943.

17. Clark, S. J . , "Quantitative Methods of Organic Microanalysis," Butterworths, Lon­

don, 1956.

18. Colson, A. F., Analyst, 77, 139 ( 1 9 5 2 ) . 19. Colson, A. F., Analyst, 80, 690 ( 1 9 5 5 ) . 20. Colson, A. F., Analyst, 83, 169 ( 1 9 5 8 ) . 21. Corning Glass Works, Corning, New York.

22. Csokan, P., Magyar Kémi. Folyoirat, 48, 56 ( 1 9 4 2 ) .

23. Cunningham, K . G., Dawson, W., and Spring, F. S., / . Chem. Soc, p. 2305 ( 1 9 5 1 ) . 24. Dimbat, M., and Stross, F. H., Anal. Chem., 29, 1517 ( 1 9 5 7 ) .

25. Forster, E. L., and Breitenbach, J . W., Mikrochim. Acta, p. 982 ( 1 9 5 6 ) . 26. Francis, H. J . , Jr., Personal communication, I 9 6 0 .

27. Furman, Ν . H., ed., "Scott's Standard Methods of Chemical Analysis," 5th ed., Vol. II, Van Nostrand, New York, 1939.

28. Gibson, D . T., and Currie, J . , Mikrochim. Acta, p. 644 ( 1 9 5 7 ) .

29. Grant, J . , "Quantitative Organic Microanalysis, Based on the Methods of Fritz Pregl," 4th ed., Blakiston, Philadelphia, Pennsylvania, 1946.

30. Grant, J . , "Quantitative Organic Microanalysis," 5th ed., Blakiston, Philadelphia, Pennsylvania, 1951.

31. Guerrant, G. O , Anal. Chem., 30, 143 ( 1 9 5 8 ) .

32. Gysel, H., and Hamberger, K., Mikrochim. Acta, p. 254 ( 1 9 5 7 ) . 33. Hawkins, J . J . , and Arthur, P., Anal. Chem., 23, 533 ( 1 9 5 1 ) . 34. Hofstaeder, R., Metropol. Microchem. S o c , New York, January I 9 6 0 . 35. Hoyer, H., Mikrochemie ver. Mikrochim. Acta, 36/37, 1169 ( 1 9 5 1 ) . 36. Iwamoto, K., Sci. Repts. Tôhoku Imp. Univ. First Ser., 17, 719 ( 1 9 2 8 ) . 37. Kamm, O., "Qualitative Organic Analysis," Wiley, New York, 1923.

38. Kofler, L., and Brandstàtter, M., Mikrochemie ver. Mikrochim. Acta, 33, 20 ( 1 9 4 8 ) . 39. Kubota, B . , and Yamane, T., Bull. Chem. Soc. Japan, 2, 209 ( 1 9 2 7 ) .

40. Magee, R. J . , and Wilson, C. L., Analyst, 73, 597 ( 1 9 4 8 ) .

(19)

2 1 . Molecular Weight 546

4 1 . Millard, Ε. B . , "Physical Chemistry for Colleges," 1st ed., McGraw-Hill, New York, 1921.

42. Milton, R. F., and Waters, W . Α., "Methods of Quantitative Microanalysis," 2nd ed., Arnold, London, 1955.

43. Morton, J . J . , Campbell, A. D., and Ma, T . S., Analyst, 78, 722 ( 1 9 5 3 ) . 44. Nash, L. K., Anal. Chem., 19, 799 ( 1 9 4 7 ) .

45. Nash, L. K., Anal. Chem., 20, 258 ( 1 9 4 8 ) . 46. Nash, L. K., Anal. Chem., 23, 1868 ( 1 9 5 1 ) . 47. Nernst, W . , Gbttinger Nachrichten, p. 75 ( 1 9 0 3 ) . 48. Neumayer, J . J . , Anal. Chim. Acta, 20, 519 ( 1 9 5 9 ) . 49. Niederl, J . B . , and Levy, A. M., Science, 92, 225 ( 1 9 4 0 ) .

50. Niederl, J . B . , and Niederl, V., "Micromethods of Quantitative Organic Elementary Analysis," Wiley, New York, 1938.

51. Niederl, J . B . , and Niederl, V., "Micromethods of Quantitative Organic Analysis,"

2nd ed., Wiley, New York, 1942.

52. Niederl, J . B . , and Sozzi, J . Α., "Microanâlisis Elemental Orgânico," Calle Arcos, Buenos Aires, 1958.

53. Nieuwenburg, C. J . van, and Ligten, J . W . L. van, Anal. Chim. Acta, 9, 66 ( 1 9 5 3 ) . 54. Ogg, C. L., / . Assoc. Offic. Agr. Chemists, 41, 294 ( 1 9 5 8 ) .

55. Ogg, C. L., Personal communication, 1 9 5 9 ; / . Assoc. Offic. Agr. Chemists, 43, 693 ( I 9 6 0 ) .

56. Pharmacopeia of the United States by the Authority of the Pharmacopeial Conven­

tion, X I I I , X I V , X V , X V I , Mack, Easton, Pennsylvania, 1947, 1950, 1955, and I 9 6 0 .

57. Pirsch, J . , Angew. Chem., 51, 73 ( 1 9 3 8 ) . 58. Pirsch, J . , Ber., 65, 862 ( 1 9 3 2 ) . 59. Pirsch, J . , Ber., 65, 1227 ( 1 9 3 2 ) . 60. Pirsch, J . , Ber., 66, 349 ( 1 9 3 3 ) . 61. Pirsch, J . , Ber., 66, 815 ( 1 9 3 3 ) . 62. Pirsch, J . , Ber., 66, 1694 ( 1 9 3 3 ) .

63. Pregl, F., "Quantitative Organic Microanalysis," ( E . Fyleman, trans. 2nd German ed.), Churchill, London, 1924.

64. Procter & Gamble Company, Cincinnati, Ohio.

65. Puddington, I. E., Can. J . Research, 27, 151 ( 1 9 4 9 ) . 66. Rast, K., Ber., 5 5 B , 1051 ( 1 9 2 2 ) .

67. Rast, K., Ber., 5 5 B , 3727 ( 1 9 2 2 ) . 68. Rieche, Α., Ber., 59, 2181 ( 1 9 2 6 ) . 69. Rieche, Α., Mikrochemie, 12, 129 ( 1 9 3 2 ) .

70. Roth, H., "Die quantitative organische Mikroanalyse von Fritz Pregl," 4th ed., Springer, Berlin, 1935.

71. Roth, H., " F . Pregl quantitative organische Mikroanalyse," 5th ed., Springer, Wien, 1947.

72. Roth, H., "Pregl-Roth quantitative organische Mikroanalyse," 7th ed., Springer, Wien, 1958.

73. Roth, H., "Quantative Organic Microanalysis of Fritz Pregl," 3rd éd., ( Ε . B . Daw, trans. 4th German ed.), Blakiston, Philadelphia, Pennsylvania, 1937.

74. Scientific Glass Apparatus Company, Bloomfield, New Jersey.

75. Signer, R., Ann., 478, 246 ( 1 9 3 0 ) .

76. Smit, W . M., Chem. Weekblad, 36, 750 ( 1 9 3 9 ) .

77. Smit, W . M., Ruyter, J . H., and Van Wijk, H. F., Anal. Chim. Acta, 22, 8 ( i 9 6 0 ) .

(20)

547 References

78. Spies, J . R., / . Am. Chem. Soc, 55, 250 ( 1 9 3 3 ) .

79. Steyermark, Al, "Quantitative Organic Microanalysis," Blakiston, Philadelphia, Penn­

sylvania, 1951.

80. Sucharda, E., and Bobranski, B . , "Halbmikromethoden zur automatischen Verbren- nung organischer Substanzen und ebullioskopische Molekulargewichtsbestimmung,"

p. 135, Vieweg, Braunschweig, Germany, 1929.

81. Sucharda, E., and Bobranski, B . , Przemyst Chem., 11, 371 ( 1 9 2 7 ) . 82. Thomas, Arthur H., Company, Philadelphia, Pennsylvania.

83. White, L. M., and Morris, R. T., Anal. Chem., 24, 1063 ( 1 9 5 2 ) .

83a. Wilson, Α., and Bini, L., Meeting-in-Miniature, North Jersey Section, American Chemical Society, South Orange, New Jersey, 1959; Anal. Chem., in press.

84. Wilson, C. L., Analyst, 73, 585 ( 1 9 4 8 ) .

85. Wilson, Η. N., and Heron, A. E., / . Soc. Chem. Ind. (London), 60, 168 ( 1 9 4 1 ) . 86. Wright, R., Analyst, 73, 387 ( 1 9 4 8 ) .

87. Zscheile, F . P., and White, J . W., Jr., bid. Eng. Chem., Anal. Ed., 12, 436 ( 1 9 4 0 ) .

Ábra

FIG. 203. Thomas-Hoover melting point apparatus, showing enlarged view of peri- peri-scopic reading device, a, and special bath assembly for molecular weight determinations  by Rast method, b
FIG. 204. Capillaries for microdetermination of molecular weight by the Rast method,  showing various stages—details of construction, dimensions are approximate
Table 36 gives the list of some possible solvents together with their melting  points, molecular melting (or freezing) point depressions and a few com­
TABLE  3 6 *  SOLVENTS
+5

Hivatkozások

KAPCSOLÓDÓ DOKUMENTUMOK

The SAMPLE COUNT column contains the number of suicide victims with rare variants in the same gene, the NUMBER OF VARIANTS column contains the number of individual variants within

Anholt (2016) says and verifies it with research results that this kind of promotion with the exception of that of the tourist destination is quite useless and has no impact,

„There is no substitute for the clinician`s standing by the ventilator, making necessary adjustments and monitoring the effects of

the diagram it can be seen that the trajectory of the constant kinetic energy motion is much smoother, the speed along the path does not fluctuate as much as in the time

In this paper we presented our tool called 4D Ariadne, which is a static debugger based on static analysis and data dependen- cies of Object Oriented programs written in

If the 95% confidence interval is calculated for the expected value from 100 different sample, than approximately 95 interval contains the true expected value out of the 100.

250 SPECTRUM ABSORPTION METHODS IN ANALYSIS OF GAS MIXTURES source tube was filled with the sample, while the absorption discharge tube was filled with light hydrogen.. Both

• used for analyzing organic molecules, usually UV light excites the sample and as part of the relaxation from the excited state the sample emits visible light, and the amount