MASS Spectrometry (MS)

Sample Ions Detector

Ionisation

m/z

EI ESI

APCI, FAB, MALDI, etc

M

+

M+H

+

M+Na

+

M+M’+H

+

M

-

fragments

Electric/magnetic field

accelerate/

circleshaped orbit

Invasive, but very sensitive (10 -15 -10 -21 mol!!)

(2)
(3)

EI (electron impact) – hard ionisation

(4)

EI

Sample holder

10-20 g Heated by program, sample holding stick

Pros

1. Most of small molecules 2. Lots of fragments

3. Programmed heating, some separation

Cons

1. M+ is small, or not observable

2. Not suitable for thermally labile compounds

200-300C

(5)

ESI (electrospray ionisation)– mild ionisation

Sample in

solution

MeOH, AcCN, H

2

O

MH

+

, MHsolv

+

, M+Na

+

Negative mode is also possible!

Sample inlet directly (in

solution) or coupled with LC

(6)
(7)

MS Spectrum

Under given circumstances the spectrum is relevant

to the molecule!

(8)

What do we expect?

C 27 H 48

M=372

(9)

Every molecule is detected separatly!!!

12

C

27

H

48

,

12

C

2613

C

1

H

48

,

12

C

2513

C

2

H

48

, etc

13

C has 1,1% abundance: I(M+1)/I(M)~N(C)

(10)

MS spectrum of chlorine containing molecule

(11)

MS spectrum of bromine containing molecule

(12)

Izotop ratios

Most abundant

Less abundant

Ratio

12

C

13

C 100 1.1 M+1

16

O

18

O 100 0.2 M+2

35

Cl

37

Cl 100 32.5 M+2

ca 3:1

79

Br

81

Br 100 98 M+2

ca 1:1

32

S

34

S 100 5 M+2

(13)

Isotope ratios if more halogens are present

Cl Cl2 Cl3

Br

Br

Br

2

3

(14)

MS spectrum of sulphur containing molecule

(15)

C H

3

CH

3

O H

3

C

CH

3

M=58!

(16)

m/z = 120 M M (%) M+1 (%) M+2 (%) C2H4N2O4 120,017107 100 3,15 0,84 C2H6N3O3 120,040916 100 3,52 0,65 C3H12N4O 120,101111 100 5,00 0,31 C4H12N2O2 120,089878 100 5,36 0,52 C6H6N3 120,056172 100 7,72 0,26

C9H12 120,093900 100 9,92 0,44

m(N)+2m(H)≠m(O) Mass defect

HRMS: High Resolution Mass Spectrum

Nominal mass

(17)

Resolution in MS

FWHM: Full width at half maximum

(18)
(19)

How much R is needed? R>30000

m/z = 120 M M (%) M+1 (%) M+2 (%)

C2H4N2O4 120,017107 100 3,15 0,84 C2H6N3O3 120,040916 100 3,52 0,65 C3H12N4O 120,101111 100 5,00 0,31 C4H12N2O2 120,089878 100 5,36 0,52 C6H6N3 120,056172 100 7,72 0,26

C9H12 120,093900 100 9,92 0,44

Nominal mass

Monoisotopic mass

(20)

Homework 3

Pipobroman is an anticancer drug. Calculate its M+, scetch its EI MS spectrum between m/z 350-360, give explanation.

Does it have characteristic IR absorption bands? If yes, which group, what wavenumber?

Hivatkozások

KAPCSOLÓDÓ DOKUMENTUMOK

In the next section, we shall prove the basic facts concerning the eigenvalues of the linear operator L under the radiation boundary conditions that shall be used in the proofs of

(A) Full mass spectrum of sunitinib and images of the distribution of the precursor molecule in tumour, liver and kidney tissues after 2 weeks of treatment.. Signal of sunitinib

In view of the potential usefulness of VOCs as markers of lung disease we developed a simple method for their measurement using Proton Transfer Reaction Mass Spectrometry (PTR-MS)

This article provides laser-ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) and secondary ionization mass spectrometry (SIMS) U-Pb and U-Th zircon dates for

The prepared chips were used for the dilution, counting and characterization of nanoparticles (NPs) performing single particle inductively coupled plasma mass spectrometry

A: A minta és referenciaanyag totálion kromatogramja. B: A minta és

The aim of this study was to identify by High-performance liquid chromatography- electrospray ionisation mass spectrometry (HPLC-ESI-MS) the main phenolic molecules

The Effect of the Plasma Sampling Depth and the Flow Rate of the Aerosol Dilution Gas on the Performance of Single Particle Inductively Coupled Plasma Mass Spectrometry..