A FAST NEUTRON TIME-OF-FLIGHT SPECTROMETER

By

G. HREHl:SS*, A. XESZ2IU'!;LYI* and K. SDIO~YI

Institute for Theoretical Electricity of the Polytechnic LniYersity, Budapest (Received ?lIarch 8. 1958)

To solve many a problem of nuclear physics, it is necessary to measure the energy of neutrons producing nuclear reactions or arising in the process of it. Such problems are, among others, the energy dependence of cross-sec- tions, especially that of the fission cross section, further the investigation of highly excited states of nuclei by means of inelastic scattering, and the mea- :3lUement of energy spectra of neutron -sources. As is well-known from literature, different apparatuses have been developed for measuring neutron energy from thermal to 10 ~IeV [1-5]. Some of these require comparatively simple equipment as e. g. the precise emulsion-, [6-7], or telescope-technique [8]

based on proton-recoil, on the other hand either their evaluation takes up much time or they possess 10"'- efficiency. Others, however, as the neutron diffractogragh or the He3 recoil method can only be applied in a narrow energy band. The time-of-flight apparatuses which are becoming more and more current, undoubtedly possess the largest energy range; the high efficiency and a relatively good resolution counterbalances their expensiveness.

Short neutron bursts at lo'w energy are usually produced by mechanical methods, in the fast region with electronical ones (by interrupting the charged particle beam which generates the neutrons). The energy can be determined hy measuring the flight time over a given distance. The quadratic character of the time-energy function and cons('quently the necessity of a redoubled accuracy in measuring time are the apparent disadvantages of this system.

On the other hand, with the lenghtening of a flight-path, if intensity alIo-ws, the time-of-flight method is applicable even in the 100-200 NIe V energy region [9].

We shall describe a fast neutron time-of-flight spectrometer for the energy range of 0,8-14 NIe Y. It 'was developed in the Department for Atomic Physics of the Central Res('arch Institute of Physics (Budapest), for the investi- gation of the reaction-mechanism of highly excited nuclei (direct interaction) hy means of inelastic scattering experiments. The gen('ral arrangement of

" At the Central Research Institute of Physics. Department for Atomic Physics.

6*

(2)

132 C. HREHUSS, A. ~YESZJH!;LYI and K. SIJfOXYI

the apparatus is similar to that of CIL4..NBERG and LEVIN [10], some details of which, however, may be of general interest.

100

20

o

For neutrons of energy E the flight-time over a distance L is :

\ I

I ! I ! i

I

I I I I

I I

te[)1 I I

L·lm i

\ I I

I

I i

I

i

I

\ I

I I I I I ! I I i I i

\ I I

I I I

I

I

I I

Dt ~ I< In).! sec

-

.'" I

i ! ,

\

~I ~

!

I

J----

I I I Llt • 2rkP sec I

...

I I

I

. /

I i ! i i

.r--

I

I

I i I i i

r

o

2 4 6 8 10 12 14 16 18

----.. £Nev

Fig. 1. Flight-time and relative energy spread as a function of nentron energy.

The flight-path is 1 III

I t

i I

!

,

-;

j

Fig. 1 sho'w5 the smallest relative errors "which can be attained at different energies, if the error of the time measurement is 2 and 4 m,usec respectively.

The reported values concern a flight path L = 1 m. Neutrons are generated in short bursts and after having been scattered they flight the path 1,2-1,8 m, with a changed energy. At the end of this they enter a fast detector. Bet"ween the pulse given by-the detector and the reference signal, appearing at another channel in the moment the deuteron bnrsts are produced, there is a time dif- ferencc: t -;- const. The two signals get into the fast coincidence unit, which gives an olitput signal only if the reference pulse is properly delayed. The delay is equal to the flight-time, which is obviously determined only to the extent of an additive constant, but it can be precisely fixed with a suitable calibration (inelastic l' rays).

To be J t = 1-3 m,usec it is necessary that the neutrons in the bursts should start with no more than the same uncertainity. This was attained as follows: a Cockroft-W alton type generator of two hUlldred kilovolts with a radiofrequency-ion source [11] supplies a well-focused D'" ion-beam of about 0,1-1,0 mA intensity. The ion-heam vertically leaving the accelerator, enters a mass and energy analyzing magnetic field, which deflects the D-+- component with 90~ to the horizontal direction (Fig. 2). The D":' -+- component

(3)

Ion Source (Radio Frequency

Type)

Acceleraflon fo 200 keV

I

-.!fiO~f'L

... _, .. _.. .15'<2. c'..'! ... _____ ._ 12cm

TODiffusion __ Pump l.

r--=1 j , [ J Cd 20

?' • - "

'~~'--' r''l

~ ~\ - =.n·~'t~' ~ = ~--~- -:,,,,,

. \~~I -=-=-==- --~I ~ll b-J

D<of{oc{~

and

~ \:~~=--;;;- -. lJoaph~gm I

ToOfffwlM

f1d"kw/y,o,,,og"'·'~i t

Pumpu'

Declrostatic Quadrupo/e ienseSystvm

(Strong focusing) lon-Bursls

P1 ~

Ior Osclllafor

Dor/ec kv ff

)6-

Un6 e

-. 81'1Hz

(-4-

Tol'1agnel Field Stabiliser

Fip;. 2. Till,,, of' I'lig:hl fast lIe"ll'''" spee\l'Olllel('l'. Gelll'raJ layollt of lh" appa"allls

ScaHerer

\ \

\

\ To Delectol

\

""

"1

""

Ch '-l '-' t>i r: ;]

o '.-:

S 8

a ';1

~

~

'-l t/,

~

~ a

~

'-l r.J

~

w

w

(4)

134 G. HREHUSS, A . • YESZ.\[EL 1'1 and K. 5/.\[0.\T1

may separately be used to stabilize the magnetic field and the accelerating yoltage, respectiyely. The deflector magnet weights about 250 kg and at 6000 gausses it requires 6 A at 300 Y. The magnet is water-cooled.

The horizontal ion beam passes between two deflector plates, the applicd r. f. voltage is max. 7 kV eff. at 4-8 :MHz. The ion-beam deyiates from the horizontal plane with a corresponding frequency. A diaphragm (-with an

---

Deflector

J)/o!?j

Slot

, I i

n n

Fig. 3. Production of very short ion-bursts

i

i i i

! I : \

, i ,

• i

d

opening of 1,5 cm, placed at a distance of I,.) m) bars the continuing route of the ions coming to the diaphragm beyond the time interyal: T/2 To

T/2

+

T. (Fig. 3.) At a giyen geometrical anangement and frcquency, ad- equately short bomharding deuteron hursts call be obtained by a suitahle yoltagc amplitude. The ions going through the diaphragm and passing through an electrostatical quadrupole lens system are focused on the target containing tritium or deuterium. This quadrupole lens system is of strong focusing and doe;;

not take part in the accelerating process. The lens requires two independent and regulated yoltages in the range of 0-5 k Y.

The vacuum system is about 7 m long (its volume being 100 1) and a prcssure of 0,8-2.10.-5 to1'1' is maintained by t·wo yaCUUlll pumps located at different parts of the system and haying a speed of 1750 ,ul/sec. The external rate of gas enter of the 'wholc systcm is negligible compared to the 10 ncm3/1t D2 gas getting into the system through thc ion-source. Along the ion-path three yacuum-mcters are placed, one of the ionization-, and two of the Penn- ing-type.

(5)

A LIST _YECTROj- TDIE-OF-FLIGHT _'PECTRWIETER 135

The intensity of the pulsed ion-beam bombarding the target is about 1 {loA and the resulting neutron yield is 106-107 neutrons/sec from the reac- tions Did, n/He3, T d, njHe4, respectively. The secondary electron current is suppressed by a reverse electric field, resulting in a leEs then one-percent error for determining the ion-current. The tritium targets are imported from abroad, while the deuterium-targets are produced at the Department [12].

12cm

Scintillation Countef' Vith Plastic Phosphor

~~i~~~~ QQd

ReA 5819 t1ulfiplier

To Fast necfronics Lead Shield/ng

Fig. -1. Physical arrangement of detector, shield and target

The scattered neutrons enter the detector through a paraffin collimator (Fig. '1) which prevents the direct target-neutronE being counted. The neutron detector is a plastic scintillator (1,5 n in diameter and 2 n long, having a decay time constant of 2,5 mflsec), which is covered, together with the ReA 5819 type multiplier, by a lead shield of ;) cm thickness. The detector and colli- mator can be rotated from 0-90: around the scattering axis to ensure the possibility for the investigation of the angular dependence of scattering (See Fig.;) and Fig. 6).

In the design of fast electronics it was necessary to insure maximum speed (Fig. 1 shows that at high energies the required frequency-band extends up to 2-300 il:IHz). A fast time-analyzer was developed consisting of the usual elements of the millimicrosecond technique. The time-measuring system is of the one-channel type. Instead of measuring the time intervals directly, we detect coincidences between the reference-pips and the neutron signals, the former being delayed in a known and variable manner. The block diagram of the electronic arrangement is sho'wn in Fig. 7. The neutron signal at the

(6)

136 G. HREHl'SS, .'1. ~l'ESZ.1It:L 'YI and K. SnIO~YYI

Fig. 5. Yic\\' of the ;201) keY accclerator, deflector magnet and high frequcney oscillator

anode of the multiplier has an amplitude of about 0) Y and a rise-time of 1-3 m/iS. After clipping with a shortened line it is amplified by a distrihuted amplifier. The characteristic data of the amplifier are the following; frequency band: 10 kHz-170 MHz, gain: 25 dB at a matched output, input imped- ance: 150 olum, tubes employed: 11 pieces of the EF 80 type [13].

The coincidence unit 'was of the crystal diode-hridge type similar to that descrihed hy }Iinton [14]. The resolution curve of the distributed ampli- fier - coincidence circuit system had a half-width of 3.10-9 sec at half- maximum, and an efficiency of 60%. It 'was checked hy direct experiments, (see Fig. 8) that the resolution time was not increased hy the rise-time limitation of the amplifier. The delay of the reference-pip was obtained hy

(7)

_-1 FAST _YECTROJ- TDIE·OF·FLICHT .'PECTROJIETER

Fig. 6. Deflector magnet, electrical deflector plates

c:cm/illator

__ ~ TDr;UTr~~r .

r----..., P." I

Idf· f80 tlHz}

las/ dynOde Crys!c1 Bridge

Coincidence Circu!/

.,..J.IO-9S

..1,-.---....,

~--~ r---~---~

S/Ol./ Amplifier and DiscrimInator

Fi::. I. In,trulllentatiflll diagram for f,j,;t neutron time-of-f1ight experiment 137

(8)

138 G. HREHCSS. A. _YESZ.UEL Y1 aad K. S1.UO_n·1

a precIsIOn widc-band phase shift er circuit, the input signal of which was derived directly from the deflector coil of the r. f. oscillator. In the 4-8 MHz band the error of thc phase shift er 'was less than 2=. The phase of the output signal could he varied continuously in the 0-360= range. The null-comparation was carried out by a symmetrically driven squaring circuit consisting of two difference-amplifier stages. After clipping "with a shortened line and further amplification the resulting signal had a half-"width of 3-4 illflsec.

The positive signal from the last dynode was used for amplitude selection, the fast coincidence output being gated by a slow system (0,08 flSCC lise- time) consisting of a slo'\\" amplifier, integral discriminator and scaler. This is necessary to prevent the counting of neutrons from previous bursts and random coincidence pulses. The output of the slow system 'was used as a monitor.

For the measurement of neutron energy spectra 'we must determine thc gated coincidence rate as the function of the phase-shift (i. e. constant time delay) "which can be directly read off on the phase shifter. As our detector

peA 5819

r -

1200 V (Ilegulafed ) I

0.l3111

,

• ; I

O?O+

I

, : I

iOnr

Sharfea Sfub , /·27cm

Jx

os

160

To Amp!ifie!'

\cJ

,---- _ u

t

I I I

1'4

Of80k

1

~

::k

Snf

!

0

i 150

I '

~

Fig. 8. Fast coincidence circuit

T

ReferenC9 Signal

has a non-negligible efficiency for y rays, the physical zero can he measured as an intense group in the neutron spectra caused hy inelastic i' rays.

To prove the whole system, we measured the direct target spectrum of a zirconium-target (Fig. 9). The double representation was introduced for control purposes: it is a direct consequence of the fact that nelitron bursts are produced at each null transition of the r. f. signal, i. e. at a double rate

(9)

A FAST J-ElTROJ- TDIE-OF-FLlGHT SPECTRO.\IETER

1.0 I

I 0.8

~ ~

~ 0.6

t:::

~ ';:;

.S:

(3 0.4

1'\1

I

n

I I

I \ I I L • 180 cm

i

\

I I I f c 411Hz

ii 1I

I

11

I ii

I L\

0.2

f

1 \

1

/ \

I I

1 \

"'- I ...- /"

I"

<p0 __

o

:';-0. 80°

4 J

120°

2

160' 200' 240' 280'

I c(

11'94 9

, , , 211410 ' . ! , ! ! 6 " ! J !

Fig. 9. Direct target spectrum of neutrons from T;d, 1l/Hc4 reaction

Fig. la. Target, collimator "'"stem and fast electronic:,

139

~

-

--

360'

(10)

140 G. HREH[-SS, A •• YESZ_UELYI and K. SIJIO.YYI

as to the reference signal. From Fig. 9 it can be seen that the spectrometer is able to inyestigate neutron spectra in the desired energy range, improve- ments of the resolution curve and hackground effects, hO'wever, are necessary and are in progress.

The physical arrangement of the target, collimator and fast electronics is shown in Fig. 10.

The authors wish to thank all those who helped them in their 'work, especially

1.

MEREY and

:;\1.

UZSOKY, for theiT excellent suggestions in the mechanical and electTonical dcsign, E. DEBRECZEj\;y and the electronic group for the development of the slow-electronics.

Summary

.:i. fast neutroll spectromcter is described, which \\"as deyeloped for the meaSU,:elllellt of inelastic neutron spectra in the energy range of 0,8-1-1 ::lIeI'. The fast 11eutrol11urst5 are produced by electronic means deflecting the 200 keY beam of a Cockroft-\'\' alto11 type accelator.

The neutron detcctor is a plastic scintilla tor ; the neutron signals are analysed by a fast time analyzer of the single.channel type. The duration of neutron hursts is 1-3 m,Usec. the time resolution is about ;:; ll1!1SCC.

References

1-3. HCGlIES, }IoSTOYOI et al; 'WLADDIIRSKY,

,rIBU);.

HAYE);s, Proc. Int. Conf. Geney"

1955, SeI. Ref. ::IIaL Yol. IY. pp 1-97.

I). ROSE);, L., STEWART, L.: Phys. Rey. 107, 82-1 (1957).

- ROSE);_ L .• Proc. Int. Conf. Genf 1955, Sel. Ref. ::IIat. Yo1. IV. p. 9,.

o. JOII);SO);. C. H., TRAIL C. C.: Re\". Sci. Instr. 27. 168 (1956).

9. SCA);LO);, J. P. et al.. Rey. Sci. Instr. 28, 7-19 (195/).

lO. CRA);BERG, L., LHI);, J. S., Phys. Rc\". 103, 3·13 (1956).

11. ERO, J., Acta PhY5ica 5, 391 (1955).

I:!.. FEllER,!', HREIIVSS, Gy., TiD cs ZrD targetek kcszitese, to he published in KFKI Kiizlemr- nyck (Proc. of the Central Research Iustitut of Physics, Budapest).

13. -:;-ESZ)IELYI, A.: Lancerositok (to he puhlished in KFKI Kozl.).

1-1. }Ir);To);, J. Res. -:;-at. Bureau of Standars 57, p. 119 (1956 Sept.).

Prof. K. SnIO:,\YI, Budapest, Budafoki lIt 8., Hungary

Hivatkozások

KAPCSOLÓDÓ DOKUMENTUMOK

The fast relaxation in the right knee extensors following maximum isometric bilateral force exertion caused the transitional, fast increase of torque in the unilateral counteracting

The first objective was to uncover the differences in the excitatory synaptic inputs of the two types of PV+ cells, the axo-axonic (AAC) and fast-spiking basket cells (FS

In this article, I discuss the need for curriculum changes in Finnish art education and how the new national cur- riculum for visual art education has tried to respond to

Mean solar time, defined in principle by the average rate of the apparent diurnal motion of the Sun, is determined in practice from a conventional relation to the observed

The final result of evaluating that node will be one (1) regardless the value of the remaining connected children nodes. See also Algorithm 4 below which describes how

Firstly, the Granger-character of the two variables is totally different – the for- mer is rather exogenous, the latter is endogenous –, secondly, energy consumption is basically

The fast increase of the subject matter in Fluid Mechanics on one hand, and the limited, in some cases even decreasing time available for teaching it, calls for the

The original GRIESS-ILOSVAY method was quite fast for qualitative test purposes; however, for quantitative analysis a rather long waiting time was necessary to