OBSERVATIONS WITH THE CA-IOOO MICROWAVE TELEMETER*

By

E. FARKAS

Research Laboratory of the Institute of Geodesy, Surveying and Photogrammetry, Technical University, Budapest

Presented by Prof. Dr. P. BIRO Received September 1st, 1976.

Since their first appearance 30 years ago, microwave telemeters have proved their suitability in geodesy. In Hungary, microwave telemeters have been applied since 15 years, the serial production of GET-Bl, stressing the relating research work. This program has been shared by the Institute of Geodesy, Surveying and Photogrammetry, experimenting with microwave equipment for years. Recent promising experiments have been concerned with telemeter types GET-A2 and CA-lOOO. Investigations made with this latter ,,;-ill be described below, together with some suggestions on its adapta- tion.

CA-lOOO manufacturer's (Tellurometer Co.) data:

Mean square error of distance measurement by CA-lOOO (15

+

5.l0-6D)mm Range: short range horn 10 km

long range horn 30 km

Measuring frequencies 19 to 25 MHz Accuracy of frequency better than 5.10-6 •

Experiments aimed at checking these data, at an eventual improvement and at the adaptation of the instrument for Hungarian conditions.

In Hungary, use of the telemeter type CA-lOOO has been suggested primarily for improving the accuracy of the geodetic horizontal control net and the economy of the determination of new control nets. In lack of a Hungar- ian test net, this was achieved, in addition to laboratory measurements, by re-measuring some distances in the existing third-order triangulation net.

Our laboratory tests concerned the inaccuracy of measuring frequencies, an important component of the errors of microwave telemeters. Frequency variation of the CA-lOOO signal vs. switched-on time, temperature and feed voltage has been investigated.

CA-lOOO telemeter is featured by an unthermostated crystal oscillator, raising doubt to the manufacturer's instruction to ignore stabilization time.

,. Abridged text of the Doctor's Thesis by the Author.

(2)

64 PARKAS

Therefore our laboratory tests involved the switching-on phenomenon at various temperatures (Fig. I). At a difference to the manufacturer's instruc- tion, neglect of the switching-on phenomenon proved to be only admissible in the temperature range from 12.5 to + 33 QC, while low (- 3 QC to 12.5 QC) and high (+33 QC to +43 QC) temperatures imposed a waiting time of 4 to 5 min before the oscillator got stabilized. At still lower temperatures, e.g.

-8 QC, the oscillator did not get stabilized even after 7 min.

, A (Hzi

, I

620

-+1----.-:

--'--'1--1'---;--" .-. =-=;;:;;O""'17"""5=--.~1 ~22-,5""'! -. 3~'C 610 t f--~+

"

~-=T,r-I;'='Te,Ol 12,0

1

I

I

I

1 - - -of' I 27,5 600-+1_-+I_'-,'~'---'-~-'~-':+: _'-_'~+~·=_~·~r3~3,~0-+1_+2_.0~

: I

.-11'. I

I !

21. S79 5S2 ~==tt=Jj,=q=s;;j;;;:;:::F=j:;:;;;;;;;;;;i;=~, 43,0 560 -!---i-/-.l-J-.---+-_+_-...;--+----'--,

570+--7T~-_+-_+_-+_--~--~

560 ,

I

r/

! I I I I 1 t ! , i

I

! , I

I,' !

r

I I

I I I

I

i

550 5£.0

530 1

o 4 7 6 (mini

Fig. 1

Feed voltage-dependent variation of the crystal oscillator frequency of the instrument was checked at different temperatures (Fig. 2), showing above 12.5 QC the instrument stabilizer to behave as well as indicated by the manufacturer, hence to be practically indifferent to feed voltage variations.

But also here, at lower temperatures (e.g. +2 QC) the feed voltage variation has to be reckoned with, it being susceptible of an error as high as I mm/km.

Temperature dependence of the measuring frequencies of telemeter CA-1000 has been tested both in dry and in humid atmosphere (Fig. 3). The unthermostated crystal oscillator was found to be fairly compensated, the maximum error of 32 Hz in the test range was within the tolerance.

The accuracy of the crystal oscillator frequency can even be improved compared to the manufacturer's data by taking its deviations at different temperatures into consideration.

Our laboratory measurement results have been compiled in Table 1.

Involving them in field measurements improves the accuracy given by the manufacturer for the instrument.

(3)

MICROTr:4 VE TELEJfETER

(H:)

-30 'C 620 -t--,----,---,--:::;;;;o.-:::c-t----::;;>'_=j 2:0

'""=====I=:::;:2:!::t:::;;;::;::=+===-j122,5

i-

12,5

7,5

600-t--+-~--r--~~---t---~

24 974 592 38,0

-l---!:::====±====:::1======:±:====~ 43,0

580~-~--~---~---

11,5 12,5 "3 13.5 IV)

Fie:. 2

630 ,

24 974 592

I I i I I I I I

-8 -3 .2 7,5 12,5 17,5 22,S 27,S 33 38 43 ('C) Fig, 3

Table I

T °C -8-0 0-4 4-10 10-12

Correction mm/km 1 0.8 0.5 0.8

T °C 24-28 28-32 32-34 43-38

Correction mm/km 0.8 0.6 0.4 0.2

5 Pcriodica Polytechnica Civil 21/1- 2

65

12-24

1 38-44

0

(4)

66 FARKAS

Mean square error In of distance measurement by microwave telemeters is much affected by the determination error of instrument constant Ak enhancing the care to be applied in AI; determination measurements, with a special consideration to the inherent errors of determining the atmospheric refraction index, of the frequency, the test control net inaccuracy, and these due to reflection. To increasc the versatility, and to deduce consequences for the adaptation, the microwave telemeter CA-IOOO has been tested in several regions of Hungary, such as in the Cserhiit Mountains, accompanied by simulta- neous measurements on the third-order trigonometric net clone by the Buda- pest Geodetic Co. using an electro-optical telemeter type AGA-6/A.

Without entering in details, from the comparison of telemeter efficiencies it appeared that while in unfavomable meteorological conditions electro- optical tclemetE'r AGA-6!A had a range as short as 2.7 km, CA-IOOO exhibited a much greater spectrum amounting to ten distances ayeraging 5 km.

Three mE'asurements on the third-order net and other three measure- ments done in Recsk and Balatonkellese under the same conditions as abovE', have been compared to those made by the electro-optical telemeters (Tahle II).

The six data gave an average instrument constant A"

=

37 mm, with a mean square error m = 15 mm, better than the manufacturer's data.

Our tests unambiguously supported the suitability of the examined microwave telemeters, as ·well as their practical use and economic efficiency.

As a matter of fact, characteristics of the type CA-IOOO corresponded to the manufacturer's data, and showed utility in Hungarian conditions. Upon adap- tation, measurement results can still be improyed. This ·work is going on,

and eyen some improvements can bE' stated (see Tahles I and TT). Fmther improvements are likely upon thc following modifications, suggested by the Institute of Geodesy, Surveying and Photogrammetry, for the next year:

a) modification of the telemE'ter antenna system, reducing the beam angle to 5 to 7° likely to reduce reflexion errors, at the same time to increase

Table IT

i

Slope distance

i

Deviation Value impro"cd

Residual

Slope distance by CA-lOOO

Station Sighting point m in:;trument dc\iation

m i mm constant mm

Third-order points CA-1000 AGA-6/A 421-12

I

324-11 2745.166 2745.124 +42 2745.129 5

316-10 I 316-11 4907.169 4907.121 +48 4907.132 11 414-14 I 4.14-02 6278.196 6278.187 + 9 6278.159 -28

RECSK EOS

109

I

105 2361.495 2361.454 EOK-2000 +41 2361.458 --;- 4

105 ! 106 1184.258 1184.226 +32 1184.221 - 5

BALATONKENESE

I

EOS

Hosszumezo

I

Serhegy 4115.310 4115.258

I

+52 4115.273 -15

(5)

JIICROWA VE TELEUETER 67

the telemeter range to at least 40 km, making it suitable for first-order geodetic net measurements;

b) creation of a test net permitting to determine the instrument con- stant at a higher accuracy;

c) construction of a tube feed line system permitting to distinguish in- herent instrumental errors from outer disturbances in laboratory condition,,;

d) in view of the appearance of programmable pocket calculators, development of programs permitting field evaluation at an increased accuracy.

(Sinusoidal approximation of the s"ing curve has already been developed for pocket calculator HP-65);

e) last but not least - ill agreement with observations made by the research team of the Karlsruhe University - determination of meteorological data prevailing at terminal", at least 10 m abovc earth surface likely to improve the reliability.

Summary

In geodetic practice, recently preference has been given to electronic geodetic instru- ments, in particular. to microwave telemeters, of them type CA-lOOO, actually manufactured by Tellurometer Co., has been investigated. Laboratory tests have been made on the dependence of the stability of crystal oscillator frequency on switched-on time, temperature and feed volt- age. The CA-IOOO telemeter has been compared to electro-optical telemeters applied on third- order control nets. Finally. some suggestions are made, likely to increase measurement accu- racy.

Dr. Ervin FARKAS, H-1521 Budapest

5*

Hivatkozások

KAPCSOLÓDÓ DOKUMENTUMOK

where

lated and can approach*almost one half. Normally, as in the case of penetrating beta rays, t h e particles do not completely lose their energy in the chamber.. sphere as collector

Repeating the whole algorithm a constant number of times can make the error probability an arbitrary small constant...

Under the hypothesis of this orthogonal factor model, the unique variances of the co- variance (correlation) matrix are presumed to be equal.. Let us suppose, that

The results show that quality of the product made by MVD combined with hot-air pre-drying is higher than hot-air dried ones.. By most of the measured quality parameters, microwave

,11, the mean error ofha:;e length caused by erroneolls assembling of tubes in seconds: the corresponding linear yalue in mm can he computed if the distance between

The navigation system has to evaluate the error of position estimation and if it exceeds a certain limit, it has to cancel it by some kind of absolute measurement.. A practical

It is also shown that in case of the investigated instruments the results on phase measurements are heing affected by not only some significant systematic error,